background image

 

 

 

 

 

 

 

 

INSTRUCTIONS FOR USE 

Preparation 

1.

 

Start by removing the cellophane wrapping from the filter cartridge. 

 

2.

 

Carefully inspect the bottom of the filter housing where the paper is looking for any tears or imperfections that 
could result in a rupture.  Any questionable defects should be addressed with AFS for a replacement prior to use, 
or use a secondary backup filter (as outlined below) 

             

 

 

Summary of Contents for L-SINGLE

Page 1: ...ter cartridge 2 Carefully inspect the bottom of the filter housing where the paper is looking for any tears or imperfections that could result in a rupture Any questionable defects should be addressed with AFS for a replacement prior to use or use a secondary backup filter as outlined below ...

Page 2: ... filter lid and place in a vacuum oven at 100 F for 30 60 minutes prior to use This helps eliminate any humidity moisture that may have accumulated in the media 4 Peel foil lid away from cartridge and discard 5 Place filter cartridge into filter housing ...

Page 3: ...pture Installation 7 a For internal systems provided by the OEM attach the filter housing to the flange using the 2 5 tri clamp The filter cartridge itself acts as the gasket Tighten bolts to 10ft lbs Place entire assembly back into column and follow reassembly procedures per manufacturers instructions b For external systems running inline assemble the base of the filter housing by installing the ...

Page 4: ...ilter housing using the filter itself as the gasket Tighten bolts to 10ft lbs Place inline assembly within the flow path of your extraction system using standard 1 5 tri clamp connections For best results control place the assembly directly below a ball valve ...

Page 5: ...dequate protection Operation 8 Operate the extraction system as normal with the fill soak process see manufacturer s recommended operating procedures 9 When it comes time for the solution to pass through the filter housing it will require Nitrogen or Argon assist do not use compressed air Set the pressure on the nitrogen regulator to no more than 60psi 10 Connect the nitrogen line to the top of th...

Page 6: ...lter 12 Once the solution is finished passing through the filter as observed by sight glasses when available you may vent off the nitrogen from the system to keep it from circulating through the system and being recovered into the recovery tank Disposal 13 Once the run is complete remove the filter cartridge from the housing assembly and discard Some contents of filters may cause irritation if han...

Page 7: ...ely into the housing sealing against the o ring Opt 2 Inline housings have a secondary tri clamp on the bottom of the housing where a secondary sintered stainless filter may be used as a backup filter in case the primary filter paper were to rupture Using a secondary process will slow the flow recovery process however add extra security in case of rupture Notice There are many factors that can aff...

Reviews: