20
Troubleshooting
EN
Unitel
8
Troubleshooting
Any malfunctions that cannot be removed by means of the measures
described in this chapter may only be repaired by the manufacturer.
Problem
Possible reason
Repair
Pointer does not move
when pump is operated
or drops back very
quickly
Connections or lines
have a leak
Seal leaking connec-
tions and lines
Tank is being filling
Measure the level after
filling
Pointer goes beyond
100 % or pump plunger
does not fully return
Measuring line clogged
or bent
Make sure there are no
bends in the measuring
line
Install a condensate
trap
Condensate trap full
Drain the condensate
trap
Measuring range not
correctly adjusted
Verify the tank height
(see chapter "Adjusting
the measuring range
and calibrating the zero
point")
Incorrect indication
Measuring range not
correctly adjusted
Verify the measuring
range (see chapter
"Adjusting the measur-
ing range and calibrat-
ing the zero point")
Zero point not correctly
adjusted
Other malfunctions
-
Contact the AFRISO
service hotline
Summary of Contents for Unitel
Page 9: ...8 Produktbeschreibung DE Unitel 4 2 Abmessungen 135 mm 175 mm 235 mm ...
Page 33: ...8 Product description EN Unitel 4 2 Dimensions 135 mm 175 mm 235 mm ...
Page 57: ...8 Description du produit FR Unitel 4 2 Dimensions 135 mm 175 mm 235 mm ...
Page 81: ...8 Descripción del producto ES Unitel 4 2 Dimensiones 135 mm 175 mm 235 mm ...
Page 105: ...8 Descrizione del prodotto IT Unitel 4 2 Dimensioni 135 mm 175 mm 235 mm ...
Page 129: ...8 Productbeschrijving NL Unitel 4 2 Afmetingen 135 mm 175 mm 235 mm ...
Page 153: ...8 Opis produktu PL Unitel 4 2 Wymiary 135 mm 175 mm 235 mm ...