6
Test Settings: General OTDR Test Modes
Setup
Parameter
General-Purpose OTDR Test Mode
Full Auto
Real Time
Expert
Wavelength
User selects single or multiple wavelength(s) at which to test.
Auto Setup
N/A
(not applicable)
[Off]:
User sets all parameters manually.
[By Range]:
User sets
[Range]
&
[Resolution]
,
[Pulse width]
&
[Averaging]
selected automatically.
Range
N/A: automatic User sets OTDR scan range.
Pulse width
N/A: automatic,
based on auto-
detected range
Automatic if
[Auto Setup]
parameter is set to
[By Range]
option.
User-set if
[Auto Setup]
parameter is set to option
[Off]
.
Averaging
N/A: 1 second
updates
Automatic if
[Auto Setup]
is set to
[By Range].
User-set if
[Auto Setup]
is set to
[Off].
Resolution
N/A: automatic User-set
[High]
or
[Normal].
Range:
When setting manually, select
[Range]
,
which is at least 20% longer than the fiber under
test.
Pulse width [Pulse]:
Select wider pulse widths to
obtain cleaner traces (less noisy). Select narrower
pulse widths to detect events which are close
together (better resolution).
Averaging Time [Averaging]:
Longer averaging
times produce cleaner traces.
Resolution:
[High]
resolution provides close data
spacing for more precise event location and better
detection of closely spaced events.
[Normal]
resolution uses more filtering to provide a cleaner
trace, but with lower resolution.