FL-100-5 Dual Fuel Level Indicator
Page 5 of 23
Aerospace Logic Inc.
3150 Ridgeway Drive, Unit #43, Mississauga, Ontario, L5L 5R5, CANADA
Tel. (905) 569-3887 | Fax. (416) 352-5854 | [email protected]
www.aerospacelogic.com
FL-100-5 Operations & Installation Manual – Ver. 2.0 – October 24, 2004
SECTION 3
Setup and Calibration
Instrument setup is only required when a parameter or display configuration is to be
changed. With proper planning you should only have to perform this function once in the
life of the instrument. All information is stored in the instruments lifetime memory which
will be retained without power for up to 100 years.
NOTE: When in the setup mode the two left most characters will have decimal
points displayed with the specific characters. The display bars as well as LEFT
and RIGHT LED indicators will be off at all times during the setup process.
STEP 1
To enter the Setup mode turn off the power to the instrument.
Then press and hold the switch in the
S
position and turn on
the power.
Do not release the
S
switch during this process.
STEP 2
When the “Set-” message is displayed release the
S
switch.
The display will now show one of the two displays as follows
depending on the instrument
b
rightness control source.