FL-100-5 Dual Fuel Level Indicator
Page 18 of 23
Aerospace Logic Inc.
3150 Ridgeway Drive, Unit #43, Mississauga, Ontario, L5L 5R5, CANADA
Tel. (905) 569-3887 | Fax. (416) 352-5854 | [email protected]
www.aerospacelogic.com
FL-100-5 Operations & Installation Manual – Ver. 2.0 – October 24, 2004
SECTION 6
Error Messages and Fault Resolution
During normal operations and the calibration process the
instrument continually monitors operations to ensure that it
operates in accordance with the necessary specifications.
While calibrating the instrument a number of parameters must
be met in order for the instrument to be commissioned in an
aircraft. This section will provide you with an identification of the
error messages as well as possible problem resolutions. In all
cases error messages are displayed as indicated to the left.
Error
Point of Occurrence Resolution
1
Reserved
Contact our Technical Support department. You should
never see this message.
2
Reserved
Contact our Technical Support department. You should
never see this message.
3
L..0
or
r..0
of the
calibration process
Cause:
The fuel sender is not providing a valid indication of fuel
for the zero fuel level. To qualify for installation into a
certified aircraft the fuel sender must provide a valid
zero fuel level indication to the instrument.
Resolution
(in order of probability)
:
1. If you have not added the unusable fuel to the
tank, do so now and return to the calibration
process.
2. The fuel sender system zero level is not
correctly set. Refer to the manufacturer
documentation and correct the problem.
3. The fuel sender system is defective. Replace it
with a new or rebuilt unit.
4. If the sender orientation is set in the
r
mode
check:
a. for an open circuit condition between the
instrument and the sender;
b. that the sender system output is not
shorted to ground.