At The Wye*
Dry Side of
the Humidifier
Remove the syringe cap from
the medication filled syringe and
attach the syringe end of the
tubing onto the syringe.
Prime the tubing until the
medication reaches the end
of the tubing.
Note: the tubing priming volume
maximum is 3.65mL
Unplug the tethered cap from
the Aerogen Solo, do not
remove it from the nebulizer.
Screw the nebulizer end
of the tubing onto the top
of the nebulizer.
Insert the syringe filled with
medication into the syringe pump.
Refer to pump manual or manufacturer
for guidance on pump usage.
Turn on continuous mode on the
controller. Press & hold the on/off
button for 3 seconds from the off
position to select continuous mode.
Note: Medication is nebulized on a drop
by drop basis. Observe continuous mode
in action.
*Parts not supplied by Aerogen. Consult manufacturer for usage guidelines.
1
2
3
4
05
Continuous Nebulization
www.aerogen.com