Aerogen
®
USB Controller System Instruction Manual
29
Sterilize Components
Steam sterilization can be performed using the following three methods:
1. Autoclave wrapped parts using steam sterilization pre-vacuum cycle,
a minimum of 134 °C (270 °F - 275 °F) for 3.5 minutes with drying cycle
(134 °C wrapped cycle).
2. Autoclave wrapped parts using steam sterilization pre-vacuum cycle,
a minimum of 121 °C (250 °F) for 20 minutes with drying cycle (121 °C
wrapped cycle).
3. Autoclave wrapped parts using steam sterilization pre-vacuum cycle,
a minimum of 134 °C (270 °F - 275 °F) for 20 minutes with drying cycle.
Note:
Sterilization using the autoclave cycle (No. 3 above) may cause
some areas of the nebulizer to become discolored.
To sterilize with hydrogen peroxide gas plasma, place wrapped
parts in a STERRAD
®
System.
Warning:
Users should refer to the product labeling for the STERRAD
®
100S Sterilization System for specific instructions regarding its correct
operation.
Prior to next use:
1. Check for cracks or damage and replace if any defects are visible.
2. Perform a functional test as described in this manual.
Summary of Contents for Aerogen USB Controller
Page 2: ......
Page 70: ...68 Aerogen This page has been intentionally left blank...
Page 71: ......