Do not use the Aerogen USB Controller adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the device should be
observed to verify normal operation in this configuration.
Portable and mobile radio frequency (“RF”) communication devices can
disrupt medical electrical equipment.
The Aerogen Solo is a single patient use device not to be used on more
than one patient to prevent cross infection.
Keep all cables tidy to avoid tripping or strangulation hazards.
Ensure that the Aerogen USB Controller cable is removed from the power
supply host using the grip feature provided.
Do not attempt to clean the device while in use.
Do not obstruct the removal of the Aerogen USB Controller AC/DC Adapter
from the mains.
Do not store the Aerogen USB Controller System in a location where it is
exposed to direct sunlight, extreme heat or cold, dust or moisture.
Do not operate the Aerogen USB Controller from USB ports on medical
equipment or non-medical equipment, the controller should only be
operated from the mains power supply.
Condensate can collect and occlude ventilator circuits. Always position
ventilator circuits so that fluid condensate drains away from the patient.
To avoid damage to the Aerogen Palladium vibrating mesh technology:
• Do not apply undue pressure to the domed aperture plate in the centre
of the nebulizer (Figure 2).
• Do not push out the Aerogen Vibronic
®
aerosol generator.
• Do not use a syringe with a needle to add medication.