Power Quality Analyzer Model 3945-B
6.5.5 Transients
The Transients window allows you to set up the criteria for capturing transients.
Figure 6-10
1. Type a name for the session, up to seven characters.
2. Select the date and time to begin and end the search for transients.
3. Select the percent deviation for voltage and current transients. The choices
available from the drop-down menu are 1, 2, 5, 10, 20, 50 and 100% of the
full scale range of measurement. See § 5.4 for detailed information on these
values.
4. Select the maximum number of transients to capture (from 1 to 50).