20
Power Quality Analyzer Model 3945-B
3.3 Mechanical Specifications
Dimensions:
9.5 x 7.0 x 2.0" (240 x 180 x 55mm)
Weight:
4.6 lb (2.1kg)
Shock and Vibration:
per EN 61010-1
Tightness:
IP 50 per EN 60529
(electrical IP2X for the terminals)
3.4 Environmental Specifications
Altitude:
Operating: 0 to 2000 meters (6560 ft)
Non-Operating: 0 to 10,000 meters (32800 ft)
Temperature and % RH:
0
10
45
75
90
95
122
95
79
68
-4
32
Temperature in °F
Reference Range
Operating Range
Storage Range
Relative humidity in % RH
3.5 Safety Specifications
Electrical Safety
600V CAT IV (probe dependent)
Pollution Degree 2
EN 61010-31: 2002
EN 61010-1: 2001
EN 61010-2-032: 1995
Electromagnetic Compatibility
Immunity: EN 61326-1+A1: 1998; IEC 61000-4-30: 2003
Emission: EN 61326-1+A1: 1998