Technical
specifications
40
100,000-count graphical multimeters
Technical specifications (cont’d)
Pulse width
PW+
PW-
MTX 3282B
MTX 3283B
Depending on triggering conditions of frequency meter.
Resolution
10 µs
Minimum pulse width
100 µs
Accuracy
0.05%
10 µs
Maximum period duration
12.5 s
Triggering threshold
20% of range except range 1000 V
AC
This threshold is positive in
, negative in
.
Additional error on measurement caused by slope when crossing zero: see §.
Measurement of duty ratio.
For negative events, cross the cables.
Clock
MTX 3281B
Relative time with reset to zero at each
powering up
MTX 3282B, MTX 3283B
Real-time clock
Accuracy
30 s / month: real-time clock deviation
Display
date - month - year / hour - minute - second
Influences
Specifications of the accuracy in the field of utilization:
Typical specifications = SPEC (1 + 0,05
t ° C)
Specification of the considered measurement. Values given p. 33 to 38.
T: difference between ambient temperature and input terminal nearest to
the field of reference.
Example :
temperature: 38° C
DC measurement: 0,02 % + 8 D
t ° C = 38 - 28 = 10° C
SPEC (38° C) = (0.02 % + 8 D) (1 + 0,05
10)
= 0.03 % + 12 D
Electric field under 3 V/m: 1 per thousand at full scale. This influence does
not depend on the temperature.
Multimeter
Traceability
Calibration
In the «
General settings
», the menu
Traceability
opens a file with
multimeter up-dates:
Model
Serial No
Firmware Version
Hardware Version
A periodic calibration (once a year) of the multimeter is necessary to guarantee
the given specification (refer to §. Maintenance, Metrological verification)
TRACEABILITY
Serial No. :
10010110
Firmware Vers. : 1.00
Hardware Vers. : C
Model No. :
3283B
Last Cal. :
27/11/2005 10:08:25
Next Cal. :
27/11/2006 10:08:25
CALIBRATION