AC/DC Current Probes MR417 & MR527
31
4.7.1 Housing Protection
Protection index:
•
IP 40 per IEC 60529
•
IK 06 per IEC 62262
Drop test per IEC/EN 61010-2-032 or BS EN 61010-2-032.
4.8 SAFETY SPECIFICATIONS
Electrical Conformity to International Standards:
This instrument is compliant with IEC/EN 61010-2-032 or BS EN 61010-2-032, 300 V
Category IV and 600 V Category III.
Double or reinforced insulation .
Type of current sensor per IEC/EN 61010-2-032 or BS EN 61010-2-032: type A .
Electromagnetic Compatibility:
The device is in conformity with standard IEC/EN 61326-1 or BS EN 61326-1.
Summary of Contents for MR417
Page 35: ...NOTES ...