AC/DC Current Probes MR417 & MR527
3.3 TURNING THE INSTRUMENT ON
To turn the clamp on, push the slide switch to the
1 mV/A
or
10 mV/A
setting:
■
MR417
1 mV/A
corresponds to the 600 A range.
10 mV/A
corresponds to the 60 A range.
■
MR527
1 mV/A
corresponds to the 1400 A range.
10 mV/A
corresponds to the 150 A range.
The green
ON
indicator should light up:
■
If the indicator blinks, less than 4 hours of battery life remains.
■
If the indicator does not light up, replace the battery (see § 5.2).
3.4 AUTO STANDBY
After 10 minutes of operation without user action (such as pressing the
DC Zero
button), the clamp automatically enters Standby mode. In this mode, the
ON
indicator
will turn OFF.
To reactivate the clamp, press the
DC Zero
button or change the switch to any
setting other than OFF.
To disable automatic Standby, press and hold down
DC Zero
when turning the
instrument ON. The ON indicator blinks to indicate that the request has been applied.
The ON indicator will show steady yellow after you release the
DC Zero
button.
3.5 DC ZERO ADJUSTMENT
NOTE:
If the measurement range (or sensitivity) is changed, the
DC Zero
must be adjusted before any further measurements are made.
The
DC Zero
must be adjusted before any series of measurements and
after each disconnection and reconnection.
To adjust the
DC Zero
:
■
With the clamp connected to the measuring instrument, select the desired
measurement range (or sensitivity) on the switch.
■
Make sure that there is no conductor in the clamp and that its jaws are closed
correctly.
■
Press the
DC Zero
button.
■
The
OL
indicator will light up for approximately three seconds to indicate that the
zero adjustment is in progress.
■
The
OL
indicator will turn off to indicate that the operation has succeeded.
■
If the
OL
indicator stays on, the zero could not be adjusted.
■
Before repeating the operation, check that the jaws closed correctly (air gaps
clean, no dust, no oxidation, etc.) and that there is no conductor in the clamp.
■
Press the
DC Zero
button again.
■ In the event of failure, or if the clamp is switched off (selector set to OFF), it is
the last adjustment of the
DC Zero
that is kept.