background image

 

99-MAN 100318.v1 09/06 

 

OPERATION 

Making Measurements with the AC Current Probe Model MN126 

• 

Connect the black (S2) and red (S1) terminals to the 2 V range of your DMM or instrument. The MN126 
has a ratio of 10:1. This means that for 10A AC in the conductor around which the probe is clamped, 1V 
AC will come out of the probe leads to your DMM or instrument. The output is 100mV AC per Amp AC. 
Select the range on your DMM or instrument which best corresponds to the measured current. If the 
magnitude is unknown, start with the highest range (2V AC) then work down until the appropriate range 
and resolution is reached. Clamp the probe around the conductor. Take the reading on the meter and 
multiply it by 10 to obtain the measured current. (e.g., 59mV reading: 59 x 10 = 590mA or 0.59A). 

Meter Reading 

10mV 120mV  1200mV 

Measured Value 

100mA = 0.1A 

1200mA = 1.2A 

12000mA = 12.0A 

• 

For best accuracy, avoid if possible, the proximity of other conductors which may create noise. 

Tips for Making Precise Measurements 

• 

When using a current probe with a meter, it is important to select the range that provides the best 
resolution. Failure to do this may result in measurement errors. 

• 

Make sure that probe jaw mating surfaces are free of dust and contamination. Contaminants cause air 
gaps between the jaws, increasing the phase shift between primary and secondary. It is very critical for 
power measurement. 

MAINTENANCE: 

Warning

 

• 

For maintenance use only original factory replacement parts. 

• 

To avoid electrical shock, do not attempt to perform any servicing unless you are qualified to do so. 

• 

To avoid electrical shock and/or damage to the instrument, do not get water or other foreign agents into 
the probe. 

Cleaning 

To ensure optimum performance, it is important to keep the probe jaw mating surfaces clean at all times. 
Failure to do so may result in error in readings. To clean the probe jaws, use very fine sand paper (fine 
600) to avoid scratching the jaw, then gently clean with a soft oiled cloth. 

REPAIR AND CALIBRATION 

You must contact our Service Center for a Customer Service Authorization number (CSA#). This will 
ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# 
on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you 
want a standard calibration, or a calibration traceable to N.I.S.T. (includes calibration certificate plus 
recorded calibration data).  

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

 

15 Faraday Drive 

 Dover, NH 03820 USA 

Tel: 

(800) 945-2362  (Ext. 360) 

 

 

(603) 749-6434  (Ext. 360) 

 

Fax: 

(603) 742-2346 or (603) 749-6309 

 [email protected] 

(Or contact your authorized distributor) 
Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available. 

NOTE: All customers must obtain a CSA# before returning any instrument. 

TECHNICAL AND SALES ASSISTANCE 

If you are experiencing any technical problems, or require any assistance with the proper use or 
application of this instrument, please call our technical hotline: 

(800) 343-1391 • (508) 698-2115 • Fax (508) 698-2118 

Chauvin Arnoux

®

, Inc. d.b.a. AEMC

®

 Instruments 

[email protected] 
www.aemc.com 

AC Current Probe 
Model MN126 

User Manual

 

 

DESCRIPTION 

The MN126 (Catalog #2129.15) is a high accuracy current probe for tight areas such as crowded 
wiring. Extends DMM AC measurements to 10A

AC

. The Model MN126 offers a 5 ft lead with safety 

4mm banana plug. 

WARNING 

These safety warnings are provided to ensure the safety of personnel and proper operation of the 
instrument. 

•  Read the instruction manual completely and follow all the safety information before attempting 

to use or service this instrument. 

•  Use caution on any circuit: Potentially high voltages and currents may be present and may 

pose a shock hazard. 

•  Read the Safety Specifications section prior to using the current probe. Never exceed the 

maximum voltage ratings given. 

•  Safety is the responsibility of the operator. 

•  ALWAYS connect the current probe to the display device before clamping the probe onto the 

sample being tested. 

•  ALWAYS inspect the instrument, probe, probe cable, and output terminals prior to use. 

Replace any defective parts immediately. 

•  NEVER use the current probe on electrical conductors rated above 300V in overvoltage 

category III (CAT III). Use only on insulated conductors. 

INTERNATIONAL ELECTRICAL SYMBOLS 

This symbol signifies that the current probe is protected by double or reinforced insulation. Use 
only factory specified replacement parts when servicing the instrument. 

This symbol signifies CAUTION! and requests that the user refer to the user manual before 
using the instrument. 

This is a type B current sensor. Do not apply around or remove from HAZARDOUS LIVE 
conductors without additional protective means (de-energizing the circuit or wearing protective 
clothing suitable for high voltage work). 

In conformity with WEEE 2002/96/EC 

RECEIVING YOUR SHIPMENT 

Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify 
your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately 
with the carrier and notify your distributor at once, giving a detailed description of any damage. 

PACKAGING 

The AC Current Probe MN126 is shipped with this instruction manual and a product warranty and 
registration card. 

INSTRUMENT COMPATIBILITY 

The Model MN126 is compatible with any voltmeter, multimeter, or other current measuring instruments 
with an input impedance of >1M

Ω

. To achieve the stated accuracy, use the MN126 with a meter having 

an accuracy of 0.75% or better. The accuracy of the meter must be added to the probe accuracy. 
 

Reviews: