2

Meaning of the symbol  

Warning !

  Please refer to the User’s Manual before using the

instrument.
In this User’s Manual, the instructions preceded by the above
symbol, should they not be carried out as shown, can result in a
physical accident or damage the instrument and the installations.

Meaning of the symbol

This device is protected by a double insulation or by a reinforced
insulation. No linking is required from the protection earth terminal
to ensure the electrical security.

Thank you for purchasing a 

C.A 5205 Multimeter

. To get the best

service from this instrument :

Q

 

 read

 this user's manual carefully,

Q

  

respect

 the safety precautions detailed.

  SAFETY PRECAUTIONS  

Q

Never use on voltage to earth circuits above 600 V, with a voltage
surge category of over III, that is to say, fixed industrial and domestic
installations (ref. IEC 664-1).

Q

To be used indoors in environments with a maximum pollution
degree of 2 (ref. IEC 664-1), a temperature between 0 and 50°C
and relative humidity below 70%.

Q

 

 

Use accessories which conform to safety standards

(IEC 1010-2-031), with a minimum voltage of 600 V and a voltage
surge of category III.

Q

Never open the multimeter case before disconnecting from any
electrical source.

Q

Never connect up to a circuit to be measured if the multimeter
case is not closed correctly.

Q

Before any measurements are taken, ensure the leads and switch
are in the correct positions.

Q

Never measure resistances on a circuit with the power switched on.

Summary of Contents for C.A 5205

Page 1: ...1 ENGLISH User manual DigitalMultimeter Model C A 5205 m V Mk V OFF HOLD RANGE HOLD RANGE MULTIMETER C A 5205 COM 600V CAT III Q Q Q Q Q MULTIMETER ...

Page 2: ...cautions detailed SAFETY PRECAUTIONS Q Never use on voltage to earth circuits above 600V with a voltage surgecategoryofoverIII thatistosay fixedindustrialanddomestic installations ref IEC 664 1 Q To be used indoors in environments with a maximum pollution degree of 2 ref IEC 664 1 a temperature between 0 and 50 C and relative humidity below 70 Q Use accessories which conform to safety standards IE...

Page 3: ...5205 is designed for the daily needs of professionals in electricity 2000 counts Automatic recognition of AC and DC Automatic or manual range selection Various measurement accessories broaden the scope of application or provide the multimeter with new functions See the list of accessories and references on the To Order page Documentation upon request Note Alwaysuseaccessoriessuitableforthevoltagea...

Page 4: ...e The 200 mV AC range is only accessible on manual Press 3 times on the RANGE button to access this range FUNCTION SELECTOR SWITCH OFF Offposition V DC and AC voltages up to 600 V ContinuitysoundtestforaresistanceR 700Ω Ω Ω Ω Ω Ω Resistances up to 20 MΩ TERMINALS Ø4mmsafetyterminals Q COM common black Q red 3 DC AND AC VOLTAGE Q Connect the leads to the multimeter and connect it in parallel to the...

Page 5: ... Ω Q Automatic range selection read the measured value Q Freeze the value if necessary by pressing the HOLD button Ω Ω Ω Ω Ω 200 Ω 2 kΩ 20 kΩ 200 kΩ 2 MΩ 20 MΩ Resolution 0 1 Ω 1 Ω 10 Ω 100 Ω 1 kΩ 10 kΩ Accuracy 1 5 rdg 1 5 rdg 3 rdg 5 counts 5counts 3 counts Voltageon 0 5 V opencircuit Protection 500 V rms and 750 V peak 5 CONTINUITY BEEPER Q Place the selector switch on the function Q Continuous...

Page 6: ...e battery LR 6 Q Low battery indication Note The voltage of the batteries is indicated approximately on the diodefunctiononopencircuit 7 3 Buzzer Q Continuous beep for the continuity test Q Intermittent beep each time the switch or buttons are pressed and for the overload indication 7 4 Climatic conditions Q Temperature Use 0 C to 50 C Storage 20 C to 60 C Q Relativehumidity Use 70 RH Storage 90 R...

Page 7: ...LEXmodel1000 24 1 1 1000AAC Cat 2112 39 Q Replacement leads Cat 2118 66 Q Soft carrying case Cat 2118 65 Q Replacement shockproof case Cat 2980 15 9 WARRANTY Our guarantee is applicable for three years after the date on which the equipment is made available copy available on request 10 MAINTENANCE For maintenance use only specified spare parts The manufacturer will not be held responsible for any ...

Page 8: ...tandards For instrument repair and or calibration please call our factory toll free at 800 945 AEMC 800 945 2362 CHAUVIN ARNOUX Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Tel 800 954 2362 603 749 6434 Fax 603 742 2346 Orcontactyourauthoriseddistributor Estimates for repairs normal recalibration and and calibration traceable to N I S T are available upon request Overseas custome...

Reviews: