7
3. SPEcIFIcAtIONS
3.1. ENvIRONMENtAL cONDItIONS
indoor use.
operating range
-10 °c to +55 °c and 10 % to 85% rh
Storage range
-40 °c to +70 °c and 10 % to 90% rh
degree of pollution
2
altitude
< 3000 m
3.2. MEcHANIcAL SPEcIFIcAtIONS
Pistol-grip Kelvin probe
dimensions grip
108 x 40 mm
probe
154 x 30 mm
thickness
28 mm
diameter of the probes
2 mm
center-to-center distance between the probes 3.5 mm
travel of the probes
7 mm
lead
3.15 m long
Spaded lugs
for diameters from 4 to 6 mm
Safety terminals
diameter of 4 mm
Mass
approximately 2 x 420 g
resistance
approximately 2 x 40 m
Ω
Straight Kelvin probe with rotation
dimensions (l x W x h)
207 x 34 x 30 mm
diameter of the probes
3 mm
center-to-center distance between the probes 8 mm
travel of the probes
10 mm
lead
3.15 m long
Spade lugs
for diameters from 4 to 6 mm
Safety terminals
diameter of 4 mm
Mass
approximately 2 x 400 g
resistance
approximately 2 x 35 m
Ω
protection index ip 40 according to iec 60 529
iK 04 according to iec 50102
drop test
according to iec 61010-1
3.3. ELEctRIcAL SAFEty
electrical safety according to iec 61010-031
Maximum current: 10a
Maximum voltage: 30V