Note that the last measurement display selected is saved when the instrument is turned 
OFF. 

 

Relative humidity (RH) 

is the ratio of the quantity of water vapor contained in a 

volume of air to the maximum quantity  this same volume of air can contain at the 
current temperature. 

 

Dew point 

is the lowest temperature to which a mass of air can be subjected, at 

constant pressure and absolute humidity,  without liquid water forming by 
saturation.  
 

HOLD Function 

Pressing the 

 

button freezes the display. A second press unfreezes it. 

 

MAX MIN Function 

You can monitor the maximum and minimum measurements by pressing the 

 button. 

This displays the words MIN MAX at the top of the display. In this mode, pressing 

 

once displays the maximum value measured during the current session. A second press 
displays the minimum. A third press restores the normal display. Subsequent presses of 

 repeat this cycle. 

 

To exit MAX MIN mode, long-press 

 

Recording Measurements 

You can start and stop a recording session on the instrument. Recorded data is stored in 
the instrument’s memory, and can be downloaded and viewed on a computer running the 

DataView Data Logger Control Panel. You can record data by pressing the 

 button: 

 

A short press (MEM) records the current measurement(s) and date/time. 

 

A long press (REC) starts the recording session. While the recording is in 
progress, the symbol REC appears at the top of the display. A second long press 

of 

 stops the recording session. Note that while the instrument is recording, a 

short press of 

 has no effect. 

 

To schedule recording sessions, and download and view recorded data, see the 
Data Logger Control Panel Help. 

 

 

 

 

Summary of Contents for 1246

Page 1: ...Thermo Hygrometer Data Logger Model 1246 Quick Start Guide ENGLISH www aemc com ...

Page 2: ...at the time of purchase or obtained by returning the instrument to our repair and calibration facility for a nominal charge The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer For recalibration please use our calibration services Refer to our repair and calibration section at www aemc com Serial Catalog 2121 73 Model 1246 Please f...

Page 3: ...This instrument must not be treated as household waste Precautions This instrument is compliant with safety standard IEC 61010 2 030 for voltages up to 5V with respect to ground Failure to observe the following safety instructions may result in electric shock fire explosion and damage to the instrument and or the installation in which it is located The operator and or the responsible authority mus...

Page 4: ...ger Control Panel by double clicking the Data Logger shortcut icon in the DataView folder placed on the desktop during the installation 6 Click Instrument in the menu bar and select Add an Instrument 7 The Add an Instrument Wizard dialog box opens This is the first of a series of screens that lead you through the instrument connection process The first screen prompts you to select the connection t...

Page 5: ...utton to enable disable Bluetooth Selecting Temperature Units Press to toggle between C and F OPERATION Making Measurements 1 Remove the cap protecting the sensor 2 If the instrument is OFF press and hold down the button until it turns ON The instrument displays the time then the relative humidity top and temperature bottom measurements The time is set via DataView 3 Press the button to change the...

Page 6: ...the maximum value measured during the current session A second press displays the minimum A third press restores the normal display Subsequent presses of repeat this cycle To exit MAX MIN mode long press Recording Measurements You can start and stop a recording session on the instrument Recorded data is stored in the instrument s memory and can be downloaded and viewed on a computer running the Da...

Page 7: ...ration traceable to N I S T includes calibration certificate plus recorded calibration data Ship To Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 800 945 2362 Ext 360 603 749 6434 Ext 360 Fax 603 742 2346 or 603 749 6309 E mail repair aemc com Or contact your authorized distributor Cost for repair standard calibration and calibration traceable to N I S T are a...

Page 8: ...01 18 99 MAN 100452 v1 Chauvin Arnoux Inc d b a AEMC Instruments 15 Faraday Drive Dover NH 03820 USA Phone 603 749 6434 Fax 603 742 2346 www aemc com ...

Reviews: