![AED CellAED User Manual Download Page 62](http://html1.mh-extra.com/html/aed/cellaed/cellaed_user-manual_2859603062.webp)
62
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (M)
RATED MAXIMUM OUTPUT
POWER OF TRANSMITTER
150 KHZ TO 80 MHZ
OUTSIDE ISM BANDS
150 KHZ TO 80 MHZ IN ISM
BANDS
80 KHZ TO 800 MHZ
800 MHZ TO 2.5 GHZ
(W)
d = [3.5/3] x
√
P
d = [12/10] x
√
P
d = [12/10] x
√
P
d = [23/10] x
√
P
0.01
0.17
0.12
0.12
.023
0.1
0.37
0.38
0.38
0.73
1
1.17
1.20
1.20
2.3
10
3.69
3.79
3.79
7.27
100
11.70
12.00
12.00
23.00
The CellAED® is intended for use in an environment in which radiated RF disturbances are controlled. The customer or the user of the CellAED® can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CellAED®
as recommended below, according to the maximum output power of the communications equipment.
10.6 RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE
CellAED® (IEC 60601-1-2:2014/EN 60601-1-2:2015 TABLE 205)