![AED CellAED User Manual Download Page 20](http://html1.mh-extra.com/html/aed/cellaed/cellaed_user-manual_2859603020.webp)
20
The CellAED® package contains:
- One CellAED® single-use automated external
defibrillator
- This is sealed inside a transparent bag and a
protective case – do not open these until the
CellAED® is needed to treat SCA
- Instructions for operating the CellAED®
6.1 UNPACKING & INSPECTING THE
CellAED®
To help ensure the integrity of the CellAED® and to
verify that it is complete and ready for use, perform
the following inspection
(Figure 1 - Figure 4B, pages
21 - 27)
:
1. Inspect
the outer packaging for signs of damage
that may have occurred during transport. If the
packaging presents rips, cracks, bends, leakages,
discolorations, or deformation, contact your
nearest distributor.
2. Open
the outer packaging and remove the
protective case. Check that the tamper evident
seal
(Figure 1 (b), page 21)
on the protective case
has not been broken. Do not remove the tamper
evident seal, and do not remove the CellAED®
from the protective case and sealed transparent
bag until required for emergency use
(Figure 1 (c)
and (d), page 21)
.
Do not store the CellAED® in the outer packaging,
as the LED indicating battery life and readiness for
use will not be visible.
3. Check
the Date of Manufacture/Use-by Date on
the back side of protective case
(Figure 4A, page
26)
. Confirm that the current date is within 13
months of the manufacturing date. The CellAED®
expires 13 months after manufacture. Verify that
the green LED indicator is blinking. The green
LED will blink every 30 seconds
(Figure 5 (e),
page 34))
, and is visible through the protective
case
(Figure 1 (d), page 21)
. A green LED blinking
means the CellAED® is ready for use.
4. Save
the outer packaging and any inserts in case
the CellAED® requires transporting in the future.
Contact RRR with any questions about the
CellAED®
(Section 12, page 69)
.
CAUTION:
Do not open the sealed tamper-proof
and protective packaging of the CellAED® unless
you intend to use the CellAED®. This packaging
protects the CellAED® from accidental activation,
and from environmental elements such as water,
heat and dust. Opening the packaging can expose
the CellAED® to environmental contaminants and
compromise its effectiveness. The CellAED® should
be disposed of safely and immediately according to
local council regulations.
Only open the sealed packaging and activate the
CellAED® prior to immediate use.
Once the unit is unsealed, it should either be used
in an appropriate manner or disposed of safely and
immediately according to local council regulations.
06