AEB G5A-80A Operation Manual Download Page 2

Product model 

The G5-

ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-

80C. 
 
Please see the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-80C User’s Guide for 
complete information on AED use.

 

 

Indications for use 

The automated external defibrillator (AED) battery supplies power to an AED as required during self maintenance, automated 
diagnoses, and defibrillation.  The G5-

ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A, 

G5A-80C, G5S-80A, and G5S-80C. 

 

Contraindications 

Automated external defibrillators should not be used when a patient is conscious or breathing normally. 

 

Battery specifications 

Battery Type: 12VDC, 7.5Ah, lithium sulfur dioxide, primary cells. 

 

Lithium Content: approximately 9.2 grams 

 

Capacity: Typically 420 shocks or 16 hours of operating time at 25°C (77°F). Actual battery operating life depends on device settings, 
usage, and environmental factors. 

 

Charge Time: Initiation of rhythm analysis to ready to shock: 15 seconds (typical); 45 seconds (maximum) With a fully charged 
battery. Initiation of rhythm analysis to ready to shock, used battery: 15 seconds typical); 45 seconds (maximum) with a batt ery that 
has been used for 15 shocks.  

 

Shelf Life (prior to insertion): A minimum of 5 years from date of refurbishing when stored from 20° to 30°C (68° to 86°F), 1 0% to 
75% RH (non-condensing). 

 

Standby Life (after insertion/between use): 4 years when operated from 20° to 30°C (68° to 86°F), 10% to 75% RH (non-
condensing). 

 

Operating Conditions: 0°C to 50°C (32° to 122°F), 0% to 95% RH (non-condensing). Lithium Content: approximately 9.2 g. 

Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-4 

 

Enclosure Protection: IP55 when housed in AED. 

 

Transportation: UN38.3 tested: T1 

– Altitude, T2 – Thermal, T3 – Vibration, T4 – Shock, T5 – External Short 

Summary of Contents for G5A-80A

Page 1: ...G5 ABE Battery For use with Cardiac Science Powerheart G5 Models G5A 80A G5A 80C G5S 80A G5S 80C Operation Manual Revision 3 4 ...

Page 2: ...Capacity Typically 420 shocks or 16 hours of operating time at 25 C 77 F Actual battery operating life depends on device settings usage and environmental factors Charge Time Initiation of rhythm analysis to ready to shock 15 seconds typical 45 seconds maximum With a fully charged battery Initiation of rhythm analysis to ready to shock used battery 15 seconds typical 45 seconds maximum with a batte...

Page 3: ... F Not tested for airborne applications Inspect battery contacts on insertion into AED and annually Clean contacts with gold contact cleaner if any sign of oxidation or corrosion is present Always keep a spare battery on hand in addition to the battery currently in operation Use one of these solutions to clean the case of the AED soapy water denatured ethanol or 91 isopropyl alcohol Use one of the...

Page 4: ...are Rescue timer Rescueprompts and CPR countdown timer NOW SAFE TO TOUCH PATIENT improperly connected to the AED cold dried ordamaged detachedfromthepatientduringarescue Function buttons Press to enter diagnostic mode or to change the prompt language Smartgauge battery status indicator ThegreenLEDsdisplaythecapacityof the battery With use the LEDs turn off as battery capacity decreases When thegre...

Page 5: ...ry history can be reviewed with AED Manager Software How to replace the battery Presstab Lift out Insert new Pressdowntohearclick Ensure RescueReady is green Note Ensure that the battery is at room temperature before inserting it into theAED AED Battery Exchange LLC 1000 Brown Street Suite 206 Wauconda IL 60084 ...