Product model
The G5-
ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-
80C.
Please see the Cardiac Science™ Powerheart® G5, models G5A-80A, G5A-80C, G5S-80A, and G5S-80C User’s Guide for
complete information on AED use.
Indications for use
The automated external defibrillator (AED) battery supplies power to an AED as required during self maintenance, automated
diagnoses, and defibrillation. The G5-
ABE is indicated for use with the Cardiac Science™ Powerheart® G5, models G5A-80A,
G5A-80C, G5S-80A, and G5S-80C.
Contraindications
Automated external defibrillators should not be used when a patient is conscious or breathing normally.
Battery specifications
Battery Type: 12VDC, 7.5Ah, lithium sulfur dioxide, primary cells.
Lithium Content: approximately 9.2 grams
Capacity: Typically 420 shocks or 16 hours of operating time at 25°C (77°F). Actual battery operating life depends on device settings,
usage, and environmental factors.
Charge Time: Initiation of rhythm analysis to ready to shock: 15 seconds (typical); 45 seconds (maximum) With a fully charged
battery. Initiation of rhythm analysis to ready to shock, used battery: 15 seconds typical); 45 seconds (maximum) with a batt ery that
has been used for 15 shocks.
Shelf Life (prior to insertion): A minimum of 5 years from date of refurbishing when stored from 20° to 30°C (68° to 86°F), 1 0% to
75% RH (non-condensing).
Standby Life (after insertion/between use): 4 years when operated from 20° to 30°C (68° to 86°F), 10% to 75% RH (non-
condensing).
Operating Conditions: 0°C to 50°C (32° to 122°F), 0% to 95% RH (non-condensing). Lithium Content: approximately 9.2 g.
Safety Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-4
Enclosure Protection: IP55 when housed in AED.
Transportation: UN38.3 tested: T1
– Altitude, T2 – Thermal, T3 – Vibration, T4 – Shock, T5 – External Short