AEB G5A-80A Operation Manual Download Page 1

 
 
 

 

 
 
 
 

 

 
 

G5-ABE Battery  

 

For use with Cardiac Science™ Powerheart® G5 

Models: G5A-80A, G5A-80C, G5S-80A, G5S-80C 

 

Operation Manual 

Revision 3.4 

Summary of Contents for G5A-80A

Page 1: ...G5 ABE Battery For use with Cardiac Science Powerheart G5 Models G5A 80A G5A 80C G5S 80A G5S 80C Operation Manual Revision 3 4 ...

Page 2: ...Capacity Typically 420 shocks or 16 hours of operating time at 25 C 77 F Actual battery operating life depends on device settings usage and environmental factors Charge Time Initiation of rhythm analysis to ready to shock 15 seconds typical 45 seconds maximum With a fully charged battery Initiation of rhythm analysis to ready to shock used battery 15 seconds typical 45 seconds maximum with a batte...

Page 3: ... F Not tested for airborne applications Inspect battery contacts on insertion into AED and annually Clean contacts with gold contact cleaner if any sign of oxidation or corrosion is present Always keep a spare battery on hand in addition to the battery currently in operation Use one of these solutions to clean the case of the AED soapy water denatured ethanol or 91 isopropyl alcohol Use one of the...

Page 4: ...are Rescue timer Rescueprompts and CPR countdown timer NOW SAFE TO TOUCH PATIENT improperly connected to the AED cold dried ordamaged detachedfromthepatientduringarescue Function buttons Press to enter diagnostic mode or to change the prompt language Smartgauge battery status indicator ThegreenLEDsdisplaythecapacityof the battery With use the LEDs turn off as battery capacity decreases When thegre...

Page 5: ...ry history can be reviewed with AED Manager Software How to replace the battery Presstab Lift out Insert new Pressdowntohearclick Ensure RescueReady is green Note Ensure that the battery is at room temperature before inserting it into theAED AED Battery Exchange LLC 1000 Brown Street Suite 206 Wauconda IL 60084 ...