FRONT AND TOP PANELS
pH ELECTRODE
CONDITIONING & MAINTENANCE
PREPARATION PROCEDURE
Remove the electrode tip protection bottle.
DO NOT BE ALARMED IF SALT DEPOSITS ARE
PRESENT. This is normal with electrodes and they will
disappear when rinsed with water.
During transport, tiny bubbles of air may form inside the
glass bulb. The electrode cannot function properly un-
der these conditions. These bubbles can be removed by
shaking down the electrode as you would do with a glass
thermometer.
If the bulb and/or junction are dry, soak the electrode in
AD6060 storage solution for at least one hour.
STORAGE PROCEDURE
To minimize clogging and ensure a quick response time,
the glass bulb and the junction should always be kept
moist.
When not in use, store the electrode with AD6060 stor-
age solution in the tip protection bottle.
NEVER USE DISTILLED OR DEIONIZED WATER
TO STORE THE ELECTRODE.