ADWA AD 130 User Manual Download Page 17

17

pH CALIBRATION

It is recommended to calibrate the instrument frequently,
especially if high accuracy is required.
The pH range should be recalibrated:
a)  Whenever the pH electrode is replaced
b)  At least once a week
c)  After testing aggressive chemicals
d)  When extreme accuracy is required

PREPARATION
The user can choose among 7 memorized buffers (pH
1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45).

Pour small quantities of selected buffer solutions into
clean beakers. For accurate calibration use two bea-
kers for each buffer solution, the first one for rinsing
the electrode and the second one for calibration.

Remove the tip protection bottle and rinse the elec-
trode with some of the buffer solution to be used for
the first calibration point.

20

Press CFM to confirm the calibration

and the meter will store the first cali-
bration point. The primary LCD will
show the calibrated reading, while the
secondary LCD will show the second
expected buffer (pH 4.01).

If a different calibration buffer is desired, select the

appropriate value with the arrow keys.

After the first calibration point is confirmed, submerge

the electrode tip approximately 4 cm into the second
buffer solution, place the temperature probe as close
as possible to the electrode and stir gently.

The LCD will flash NOT READY for 12 seconds.

Then, if the reading is not close to the selected buffer,
WRONG BUFFER and WRONG ELECTRODE
will blink alternatively; if it is close to the selected
buffer, the calibration is performed and the display
will show READY and blinking CFM.

Summary of Contents for AD 130

Page 1: ...mV Temperature Portable Meter USER MANUAL w w w a d w a i n s t r u m e n t s c o m MANAD130R2 05 09 ADWA HUNGARY Kft Als kik t sor 11 6726 Szeged Hungary Tel 3662317878 Fax 3662550610 www adwainstru...

Page 2: ...d carefully this manual before starting operations This instrument is in compliance with the EMC Directive 89 336 EEC and Low Voltage Directive 73 23 EEC for electrical equipments For additional techn...

Page 3: ...ur dealer or the nearestAdwa Office for the RGA Return Goods Authorization number to put on the outside of your package Warranted service will be made without charge The meter is warranted for a perio...

Page 4: ...on up to 0 1 mV SYMPTOM Display shows blinking battery symbol Meter does not work with temperature probe Meter fails to calibrate or gives faulty readings WRONG CAL is displayed during pH calibration...

Page 5: ...e excessive drift Reading fluctuate up and down noise The display shows blinking full scale value mV scale out of range Display shows blinking C or F PROBLEM Dirty pH electrode Clogged dirty junction...

Page 6: ...ntinuous use Auto off User selectable 20 minutes or disabled ELECTRODES PROBES AND SOLUTIONS AD2005B 1 pH electrode with Epoxy body tip pro tection bottle BNC connector and 1 m cable AD2015B 1 ORPelec...

Page 7: ...racks on the electrode stem or bulb If any scratches or cracks are present replace the electrode Rinse off any salt deposits with water Connectors must be perfectly clean and dry CLEANING PROCEDURE So...

Page 8: ...ot function properly un der these conditions These bubbles can be removed by shaking down the electrode as you would do with a glass thermometer If the bulb and or junction are dry soak the electrode...

Page 9: ...buffer value 8 CAL key to enter exit calibration mode 9 RANGE key to select pH or mV reading 10 Secondary display 11 Primary display 12 BNC connector for pH or ORP electrodes 28 TEMPERATURE COMPENSAT...

Page 10: ...o utilize the meter sATC capability but it can also be used independently to take temperature measurements If the probe is disconnected the tem perature can be set manually using the arrow keys see pa...

Page 11: ...ution in the tip protection bottle To save battery life the auto off feature turns the meter off after 20 minutes with no button pressed This feature can also be disabled see Setup Menu section for de...

Page 12: ...sec onds and then the meter turns off pH MEASUREMENTS Remove the electrode tip protection bottle Submerge the pH electrode tip 4 cm and the temperature probe into the sample to be tested If necessary...

Page 13: ...ic temperature compensa tion ATC capability submerge the temperature probe into the sample as close to the electrode as possible and wait for a few minutes If manual temperature compensation MTC is de...

Page 14: ...9 9 mV range are displayed with 0 1 mV resolution while outside this range the reso lution automatically switches to 1 mV 23 SETUP MENU The instrument allows the user to configure the param eters thro...

Page 15: ...To avoid erroneous procedures the meter automati cally skips the buffer used for the first calibration point A difference of at least 1 5 pH units is required be tween the two buffers used for offset...

Page 16: ...stored in the instrument memory and can be recalled at any time by pressing the MRCL key The H symbol blinks to indicate that the display is in hold mode Press the HOLD key again to return to normal...

Page 17: ...th some of the buffer solution to be used for the first calibration point 20 Press CFM to confirm the calibration and the meter will store the first cali bration point The primary LCD will show the ca...

Page 18: ...the selected buffer WRONG BUFFER and WRONG ELEC TRODE will blink alternatively if the reading is close to the selected buffer the calibration is performed and the display will show READY and blinking...

Reviews: