R3765/67G Series Network Analyzer Operation Manual
10.8 Crosstalk
10-45
(11) Crosstalk of TEST PORT 4 (When the measurement path is S42) (OPT14)
Testing procedure
Setup the R3765CG/67CG as follows.
Figure 10-27 Crosstalk of TEST PORT 4 (When the measurement path is S42)
Setup the R3765CG/67CG.
MEAS
: S42
RBW
: 100Hz
Average : 16 times
Connect the short standard to TEST PORT 2.
Connect the load standard to TEST PORT 4.
Read the crosstalk value of TEST PORT 4 from the trace data.
<Check> :
Crosstalk of TEST PORT 4
300kHz to 40MHz : -90dB or less
40MHz to 2.6GHz : -95dB or less
2.6GHz to 3.8GHz : -85dB or less
3.8GHz to 5.0GHz : -75dB or less (R3767CG only)
5.0GHz to 8.0GHz : -65dB or less (R3767CG only)