![Advantech POC127 User Manual Download Page 8](http://html1.mh-extra.com/html/advantech/poc127/poc127_user-manual_2857562008.webp)
POC-127 User Manual
viii
Information on potential electromagnetic interference and advice on
how to avoid or minimize such interference (IEC/EN 60601-1-2:2007)
Guidance and manufacturer’s declaration - electromagnetic immunity
The model POC-127 is intended for use in the electromagnetic environment specified below.
The customer or the user of the model POC-127 should assure that it is used in such an
environment.
Immunity test IEC 60601
test level
Compli-
ance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2,5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
model POC-127, including cables, than the rec-
ommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1,2
d = 1,2
80 MHz to 800 MHz
d = 2,3
800 MHz to 2,5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the recom-
mended separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equip-
ment marked with the following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be consid-
ered. If the measured field strength in the location in which the model POC-127 is used
exceeds the applicable RF compliance level above, the model POC-127 should be observed
to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the model POC-127.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Summary of Contents for POC127
Page 1: ...User Manual POC 127 Intel Pineview M N450 CPU based Point of Care Terminal with 12 1 LED panel...
Page 13: ...Chapter 1 1 General Information...
Page 18: ...POC 127 User Manual 6...
Page 19: ...Chapter 2 2 System Setup...
Page 26: ...POC 127 User Manual 14...
Page 27: ...Chapter 3 3 Chipset...
Page 32: ...POC 127 User Manual 20 Step 4 Read file information and click Next to proceed...
Page 38: ...POC 127 User Manual 26...
Page 39: ...Chapter 4 4 Audio Interface...
Page 43: ...Chapter 5 5 PCI Express Ethernet...
Page 48: ...POC 127 User Manual 36...
Page 49: ...Chapter 6 6 Touch Panel...
Page 57: ...Chapter 7 7 Optional Devices...
Page 76: ...POC 127 User Manual 64...
Page 77: ...Chapter 8 8 Utility and Hot fix...
Page 79: ...Chapter 9 9 Operation and Safety information...
Page 81: ...Appendix A A MB Connector Map and Table...
Page 84: ...POC 127 User Manual 72...
Page 85: ...Appendix B B PCM 8708 MB Jumper setting...
Page 88: ...POC 127 User Manual 76...
Page 89: ...Appendix C C POC 127 Cleaning and Disinfecting...
Page 91: ...79 XXX XXXX User Manual Appendix C POC 127 Cleaning and Disinfecting...