Figure 2-5: PDC-W240 Equipotential Terminal Pin
Step 2. Prepare the Grounding cable and the other terminal link to hospital
ground/earth system.
Figure 2-6: Grounding cable with connector
Step 3. Grounding cable plug with PDC-W240 Equipotential Terminal (See Figure
2-5)
Contact your distributor, sales representative, or Advantech's customer service cen-
ter for technical support if you need additional assistance. Please have the following
information ready before you call:
Product name and serial number
Description of your peripheral attachments
Description of your software (operating system, version, application software,
etc.)
A complete description of the problem
The exact wording of any error messages
Symptoms, photo or video if available.
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The model PDC-W240 is intended for use in an electromagnetic environment as specified
below. The customer or the user of the PDC-W240 should assure that it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environmen-
tal Guidance
The model PDC-W240 uses RF
RF emissions
energy only for its internal func-
Equipotential Terminal