HIT-W183 User Manual
16
Guidance and manufacturer’s declaration – electromagnetic immunity
The HIT-W183 terminal is intended for use in the electromagnetic environment specified
below. Customers or users of the HIT-W183 terminal should ensure the device is operated
in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environ-
ment –guidance
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, con-
crete, or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for power
supply lines
1 kV for input/out-
put lines
2 kV for pow-
ersupply lines
1 kV for input/
output lines
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Surge
IEC 61000-4-5
1 kV line(s) to
line(s)
2 kV line(s) to
earth
1 kV line(s) to
line(s)
2 kV line(s) to
earth
Mains power quality should
be that of a typical commer-
cial or hospital environment.
Interruptions and
voltage varia-
tions on power
supply input lines
IEC 61000-4-11
<5%
U
T (>95% dip
in
U
T) for 0,5 cycle
40%
U
T
(60% dip in
U
T)
for 5 cycles
70%
U
T
(30% dip in
U
T)
for 25 cycles
<5%
U
T
(>95% dip in
U
T)
for 5 seconds
<5%
U
T
(>95% dip in
U
T) for 0,5
cycle
40%
U
T (60%
dip in
U
T) for 5
cycles
70%
U
T (30%
dip in
U
T) for
25 cycles
<5%
U
T
(>95% dip in
U
T) for 5 sec-
onds
Mains power quality should
be that of a typical commer-
cial or hospital environment. If
users of the HIT-W183 termi-
nal require continued opera-
tion during power
mains interruptions, it is rec-
ommended that the HIT-
W183 terminal be powered
from an uninterruptible power
supply or battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commer-
cial or hospital environment.
NOTE:
U
T is the AC mains voltage prior to application of the test level.
Summary of Contents for HIT-W183
Page 1: ...User Manual HIT W183 Healthcare Infotainment Terminal Computer...
Page 8: ...HIT W183 User Manual viii...
Page 19: ...11 HIT W183 User Manual Chapter 2 System Setup 3 Open the cover to access the RJ12 connector...
Page 26: ...HIT W183 User Manual 18...
Page 27: ...Chapter 3 3 Operation and Safety...
Page 30: ...HIT W183 User Manual 22...
Page 31: ...Appendix A A Watchdog API and Audio Path...
Page 36: ...HIT W183 User Manual 28...
Page 37: ...29 HIT W183 User Manual Appendix A Watchdog API and Audio Path...