
OsmoTECH XT Single-Sample Micro-Osmometer Service Guide
52
Appendix D:
Known limitations
•
Additives such as anticoagulants from collection tubes may contribute to the measured osmolality.
•
Particulate matter can cause premature crystallization, such as sample pre-freeze. The particulate matter may be
eliminated by centrifugation. It may also be eliminated in urine by filtration.
•
Samples with air bubbles do not freeze reliably and may yield inaccurate results.
•
Samples allowed to evaporate will yield higher osmolality values.
•
Non-aqueous solutions will not yield accurate results. The OsmoTECH XT is intended to test aqueous solutions only.
•
Highly viscous solutions may not freeze reliably and may yield inaccurate results.
•
High concentrations of ethanol, acetone, methanol, isopropanol, ethylene glycol, diethyl ether, paraldehyde,
trichloroethane, or propylene glycol may not freeze reliably and may yield inaccurate results.