Cleaning the instrument exterior : Periodically wipe the instrument with a barely-damp cloth. User either warm soapy water
or isopropyl alcohol to dampen the cloth.
•
Inspect sample probe daily
•
If sample probe shows sign of fouling, it must be cleaned by gently wiping with a lint-free cloth dampened with isopropyl alcohol or
cleaning solution for protein removal, such as Thermo Scientific Orion Cleaning Solution 900021.
•
Why?
A dirty sample probe may affect accuracy and repeatability.
•
Clean cooling well weekly
•
Use Swab Cleaner Kit dampened with isopropyl alcohol to clean cooling well.
•
Why?
Buildup in cooling well may affect accuracy and repeatability and may cause “Sample Did Not Freeze” errors.
•
Examine exhaust fans monthly
•
Ensure exhaust fans are unobstructed by dust or debris.
•
Why?
Dirty exhaust fans will cause internal instrument temperature to rise, which may cause test errors or component failures.
•
Cleaning the Solenoid Semiannual
•
Instruments used daily should be cleaned monthly, while instruments that are used less frequently should be cleaned every three months.
•
Why?
A dirty solenoid can cause “Sample Did Not Freeze” errors and can affect instrument accuracy and repeatability.
Maintenance
Refer to User Guide for more information.
Summary of Contents for OsmoTECH PRO
Page 1: ...Installation Training Guide OsmoTECH PRO Osmometer MP00082 Rev 0 PCN01026 ...
Page 12: ... Major components Instrument Operation ...
Page 13: ... Touchscreen Home Screen Instrument Operation ...
Page 14: ... Touchscreen Legend Information Instrument Operation ...
Page 16: ... Daily procedure Instrument Operation ...
Page 38: ...Accesory 38 EPSON Dot matric printer 135022 Dot Matrix Printer Paper AN2TP5 ...