Advanced Instruments Inc.
by purging with zero or a gas with a low PPM oxygen concentration is recommended after the cleaning process is
completed.
Mounting:
The analyzer is approved for indoor as well as outdoor use. However, avoid using the analyzer in an
area where direct sun might heat up the analyzer beyond the recommended operating temperature range.
Gas Connections:
The Inlet and outlet vent gas lines require 1/8” or ¼” stainless steel compression type tube
fittings. The sample inlet tubing must be metallic, preferably SS. The sample vent line may be of SS or hard plastic
tubing with low gas permeability.
Power:
The analyzer is powered by an integral lead-acid rechargeable battery. The analyzer will continue to run for
a minimum of 30-60 days after the battery is fully charged without the pump. If the pump is used, the battery will
continue to power the pump for up to 8 hours.
WARRNING: THE ANALYZER BATTERY MUST BE CHARGED IN A SAFE AREA ONLY BY USING FACTORY PROVIDED
WALL PLUG-IN CHARGER.
8