Advanced Instruments Inc.
Default Span
The software will set the SPAN adjustment based on the average output of the oxygen at a specific oxygen
concentration. For example, with factory default settings, when a span gas is introduced, the micro-processor will
display oxygen reading 50% of the span gas value, indicating that the sensor output is within the specified
limits. This feature allows the user to check the sensor’s signal output without removing it from the sensor housing.
1.
Access the MAIN MENU by pressing the MENU key.
2.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3.
Press the ENTER key to select the highlighted menu option.
The following display appears:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
4.
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT SPAN.
5.
Press the ENTER key to select the highlighted menu option.
The following displays appear and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
SET
0.1 PPM
AUTO SAMPLING
10 PPM RANGE
76 F 100 KPA
Analog Output Adjustment at known O
2
In rare instances the 0-1V signal output may not agree to the reading displayed by the LCD. This feature enables the
user to adjust the 0-1V signal output should the LCD display not agree. Note: Adjust the 0.0V signal output with the
OUTPUT ZERO option described above.
1.
Access the MAIN MENU by pressing the MENU key.
2.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
3.
Press the ENTER key to select the highlighted menu option.
The following displays appear:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
26