![Advanced Instruments 3320 User Manual Download Page 63](http://html.mh-extra.com/html/advanced-instruments/3320/3320_user-manual_2856031063.webp)
43
Troubleshooting Table
Error message:
“
Sample Did Not
Freeze
”; Impact does
not
occur
Error message:
“
Standards Reversed?
Please Repeat...
”
Error message:
“
T E Driver Failure
”
Error message:
“
Test Time-out
”
Error message:
“
Low Battery
”
Error message:
“
Cooling System
Error
”
Solenoid impactor may need cleaning. Refer to
solenoid cleaning instructions in Chapter 5. If
the error persists, contact Advanced Instruments
Hot-Line Service.
This message will appear during the calibration
procedure if the instrument detects that the low
and high calibration standards have been intro-
duced in the wrong sequence. Retry the calibra-
tion, being sure to follow the displayed prompts.
This message indicates a problem with the ther-
moelectric cooling module. Restart the instru-
ment.
The instrument was unable to complete the test
within the allotted time.
• Confirm good technique and sampler condi-
tion.
• Assure that sample probe has been fully
inserted.
This message indicates that the internal battery
needs replacing. This part cannot be serviced by
the user. Contact Advanced Instruments Hot-
Line Service.
This error can be produced in two ways:
1) When the cooling chamber is below 0ºC
before diagnostics starts;
2) When the block channel fails to get below
0ºC during the cooldown portion of diagnos-
tics, but the sample channel does.
Try restarting the instrument. If the error per-
sists, contact Advanced Instruments Hot-Line
Service.
Problem/Message
Explanation
Summary of Contents for 3320
Page 1: ...The Advanced Micro Osmometer Model 3320 User s Guide 3325 Rev12 041417 ...
Page 6: ...vi Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 8: ...viii Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 26: ...6 Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 42: ...Advanced Micro Osmometer Model 3320 User s Guide 22 Notes ...
Page 74: ...Advanced Micro Osmometer Model 3320 User s Guide 54 Notes ...
Page 80: ...Advanced Micro Osmometer Model 3320 User s Guide 60 Notes ...
Page 90: ...Advanced Micro Osmometer Model 3320 User s Guide 70 Date Problem Symptom Action ...