Error message:
“
Parameter RAM
Failed” or “No
Parameters in RAM
”
Error message:
“
New Software
Version
”
Error message:
“
Recalibration
Needed
”
Error message:
“
Sample Pre-freeze...
”
Error message:
“
Sample Probe Open/
Block Probe Open
”
Error message:
“
Sample Did Not
Freeze
”; Impact does
occur
This message indicates that the information
stored in parameter RAM has been cor-
rupted. Restore probe bin numbers, date,
time, and any other custom settings.
This message indicates that a new software
version has been installed.
This message indicates a need to recalibrate
the instrument, and will normally appear
after the installation of new software, or
when probe bin numbers have changed.
Recalibrate, following the instructions in
Chapter 3.
A sample pre-freeze message usually
appears when the sample freezes premature-
ly.
• Confirm good technique and sampler con-
dition.
• Check for particulate matter in the sam-
ple.
• Check probe bin numbers.
• Possible sample probe problem.
Check the sample probe by running the A/D
tests (see
Using the Diagnostics Menu
,
Chapter 5).
Sample may be above range of instrument.
• Run Clinitrol™ Control.
• Check probe bin numbers.
• Confirm good technique and sampler con-
dition.
Problem/Message
Explanation
Summary of Contents for 3320
Page 1: ...The Advanced Micro Osmometer Model 3320 User s Guide 3325 Rev12 041417 ...
Page 6: ...vi Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 8: ...viii Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 26: ...6 Advanced Micro Osmometer Model 3320 User s Guide Notes ...
Page 42: ...Advanced Micro Osmometer Model 3320 User s Guide 22 Notes ...
Page 74: ...Advanced Micro Osmometer Model 3320 User s Guide 54 Notes ...
Page 80: ...Advanced Micro Osmometer Model 3320 User s Guide 60 Notes ...
Page 90: ...Advanced Micro Osmometer Model 3320 User s Guide 70 Date Problem Symptom Action ...