44
i
mmuNity
ieC
60601
t
eSt
l
eVel
C
om
-
PliaNCe
l
eVel
ª
e
leCtromagNetiC
e
NViroNmeNt
-
g
uidaNCe
Electro-
static
discharge
(ESD) IEC
61000-
4-2
± 8 kV
contact ±
15 kV air
± 8 kV
contact
± 15 kV
air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%. As with the operation
of other electronic devices,
precaution should be taken to
not generate ESD.
Power
frequency
(50/60
Hz)
Magnetic
Field IEC
61000-
4-8
30 A/m
30 A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.