background image

6.0  CONFIGURATION 

All ADTE CH units are factory calibrated per P.O. instructions. 

Usually, a complete recalibration is not required unless you 

want to change input type, output type, or the range of the 

unit. A calibration sticker located on the unit identifies the 

model, calibration and options present. If recalibration to the 

same range is required, follow section 6. 1. 

For new range: 

A. Open the case to gain access to the unit's pc boards.

The larger pc board is the input pcb and the smaller pc

board is the outpuVpower.

B. To determine the MAJOR RANGE setting calculate:

1. Input Span (mV) = millivolts full scale - millivolts

zero scale.

2. Input Zero (mV) = millivolts zero scale.

3.  The Major range is the number from Table 1 that is

just larger  (in magnitude) of the  two calculations

above. Set the jumper in Table 1 to the major range

determined.

C. If the input range includes readings below 0

°

C (32

°

F)

you must set the Input Zero Control for 20 mV elevation,

otherwise set it to 0mV elevation. See Table 2 for the

jumper setting.

D. Select the TC burnout action from Table 3.

E. Select the TC type from Table 4.

6.1  CALIBRATION 

To perform a calibration check or re-calibration of the instru­

ment, follow this procedure. 

A. Make sure the unit I / 0 wiring is properly connected

and that the correct power source per the label is also

connected. The instrument must be at normal power for

a minimum of 2 minutes before proceeding to B.

B.  The input signal source must be adjustable from 0 to

100% in steps of  10% or at least 25%.  The source

should  be either precalibrated or an accurate meter

must be  used to  monitor the input. The input signal

needs to conform to an acceptable method of simulating

a thermocouple signal.

C. The  output  may  be  monitored  as  a  current  or as a

voltage across a resistor shunt,  e.g.  1-5  vdc across

250 ohms.

,�, 

MILLIVOLT 

SOURCE 

(�) 

(�\ 

DVM 

D. Apply  an  input  corresponding  to  the  zero  input

temperature.  Rotate  the 

FINE 

ZERO  control  fully

counterclockwise.  Rotate  the 

COARSE 

ZERO

switch(SWI) to a position where the zero level just goes

above  4.00 ma  output.  Rotate  the 

COARSE 

ZERO

switch back one number less than the previous position

(but not less than zero).

E. With the  input set  at the  zero input,  adjust  the 

fine

ZERO control for 4.00 ma to the desired accuracy.

F. Apply  an input  corresponding  to  the  full  scale  input

temperature.  Turn  the 

FINE 

SPAN  control  fully

counterclockwise.  Turn  the 

COURSE 

SPAN

switch(SW2) to a position where the output just exceeds

20.00 ma output. Tum the switch back one number less

(but not less than 0).

G. With the input set to the full scale input, adjust the fine

span control for 20.00 ma to the desired accuracy.

H. Repeat steps E and G until the readings remain within

to desired calibration accuracy.

I. Check the instrument at the 25-50-75% input settings

minimum.

7.0  FIELD TROUBLE SHOOTING GUIDE 

This section offers a simple, first level trouble-shooting aid 

for an apparent instrument malfunction. 

SYMPTOM 

No output 

CORRECTIVE ACTION 

1 . Check the input and output connections 

carefully. 

2. Check that the power supply polarity is 

correct and that the output loop power is

present on the indicated terminals.

3. Check  that  the  input  source  is  correct

and that it changes magnitude between

zero  and  full  scale  values  when  so

adjusted.

4. Make sure the  output loop is  complete

and  that  the  correct  meter  range  is

selected.

All external  checks are  complete. Problem 

seems to be internal. 

Troubleshooting beyond the above may be difficult without 

special equipment. We do not recommend attempting repair 

of the unit in the field.  ADTECH offers a very responsive 

repair policy. 

□□□□□□□

�  '?-

0  D 

01�1 

0  ©@©  0 

-' 

:� 

SHUNT 

(+)  2SO��m  (-) 

POWER 

SUPPLY 

.... 

....�--, 

.... 

(  : 

,""·, 

COPPER_/ 

I  I 

l

@1N 

OUT

:� 

0.01o/. 

REF. JUNCTION 

WIRE 

:_  - _  !  \��f

(ICE BATH OR ELECTRONIC) 

TC EXTENSION WIRE 

Typical Test Connection 

-2-

(9/96)  190-A-000077-A 

Summary of Contents for TCX 226

Page 1: ...llation area loca tion must agree with the supplied instruments including operating temperature and ambient conditions For detailed mounting and installation information refer to page 4 inside Electrical Connections The wire used to connect the instrument to the control system Input Output should be twisted pair s and sized according to normal practice Shielded cable is not normally necessary if u...

Page 2: ... g 1 5 vdc across 250 ohms MILLIVOLT SOURCE DVM D Apply an input corresponding to the zero input temperature Rotate the FINE ZERO control fully counterclockwise Rotate the COARSE ZERO switch SWI to apositionwhere the zero leveljust goes above 4 00 ma output Rotate the COARSE ZERO switchbackone numberless thanthepreviousposition but not less than zero E With the input set at the zero input adjust t...

Page 3: ...d J2 C TABLE 2 INPUT ZERO ELEVATION CONTROL 20mV J3 A 0mV J3 B Jumper J6 A Fixed TABLE 3 TC BURNOUT ACTION Up Scale JS A Down Scale J5 B E J T K R S B TABLE 4 TC TYPE J1 A J1 B J1 C J1 D J1 Out 8 1 PCB LAYOUT COARSE SPAN COARSEZERO FINE ZERO FINE SPAN TB1 P1 G G I I 3 9 96 190 A 000077 A ...

Page 4: ...F 31 to 85 C operating 40 to 200 F 40 to 93 C storage k Power Supply Effect 0 005 over operating range m Isolation Input output case 1000 vdc or 600 vac n Cold Junction Compensation Error 1 5 C max 0 to 50 C o Burnout Current 0 1 micro amperes nominal Note All accuracies are given as a percentage of span POWER a 8 to 42 vdc standard MECHANICAL a Electrical Classification General purpose b Connecti...

Reviews: