6-6
ADS ECHO Manual
Battery After Removal
5.
Connect the new battery and verify that a green LED light illumines. The green LED will be solid
once the
ECHO
is powered on.
6.
Place the top part of the
ECHO
onto the bottom part, ensuring that the curved back (non-sensor end)
of the top matches with the curved back of the
ECHO
bottom.
7.
Tighten the screws beginning with the middle screws on either side of the unit and tighten, not
completely, until secure. Then tighten, again not completely, the remaining screws in an alternating
sequence (for example: front left, back right, front right, back left). Once all screws are secure, fully
tighten to complete reassembly. Tighten until they no longer turn but be sure not to over tighten. Stop
tightening once there is any resistance when turning the screw.
Note:
Opening the cover to the monitor enclosure during battery pack replacement increases the
risk of damaging or introducing debris or moisture to the circuit board. Make every effort to
prevent water, dirt, and debris from contacting the monitor's circuit board during routine
maintenance.
8.
Call the monitor or connect to the monitor on site using a Bluetooth connection cable to verify
communication.
Replacing the SIM Card in the Monitor
The
ECHO
monitor performs wireless communications through an internal modem that is supported by a
SIM card located inside the monitor enclosure in a holder on the PCB. The SIM card typically requires
replacement only upon failure, which is rare, or when a change in a wireless carrier occurs. The
ECHO
comes with a SIM card pre-installed. If a SIM card not supplied by ADS is to be used, the installed SIM
card will need to be replaced.
Certain static IPs may be used for communicating wirelessly through the
QstartXML
™
software platform
or networks supporting ADS hosted system software. The service must support Mobile Terminated service
as well as Mobile-Originated (for alarm cry out) functionality. Communication with an
ECHO
requires
entering the IP address for the location in the
QstartXML
software.
When using a static IP with a
public
IP address, the PC must have internet connectivity. When using a
static IP with a
private
IP address, the PC first must establish a VPN connection with the provider. (Please
consult ADS Client Services for more information on using private addresses.) Then, the user can connect
to the monitor through
QstartXML
, provided the location has been properly configured.
Summary of Contents for ECHO 9000
Page 13: ...Introduction and Overview 1 9 First Page of the ECHO IECEx Certificate of Conformity ...
Page 16: ...1 12 ADS ECHO Manual Sample ECHO Declaration of Conformity Page 1 ...
Page 17: ...Introduction and Overview 1 13 Sample ECHO Declaration of Conformity Page 2 ...
Page 61: ...Hardware Installation 3 21 Manhole Depth and Physical Offset Measurements ...