<3. Installation>
21
IM 01E25A01-01EN
(4) Rotate the transmitter in the desired direction.
The direction can be changed to -90 degrees, +90
degrees or +180 degrees.
NOTE
The transmitter and the terminal box can be rotated -90
degrees to +180 degrees at every 90 degrees from the
arrow mark indicating the flow direction. Do not exceed
these angle.
(5) Tighten diagonally the four hexagonal bolts with the
hexagonal wrench, and then be sure to check that the
transmitter was firmly fixed to the sensor.
F0324.ai
AXW Integral Flowmeter
±0 degrees
+90 degrees
+180 degrees
-90 degrees
Upstream
side
Downstream
side
3.5.2 Remote Sensor
IMPORTANT
The following types can not be changed direction of
electrical connection after delivery.
• Submersible Type (Model: AXW
W)
• For District Heating and Cooling or Condensation-
proof (Optional code: DHC)
(1) The following tools are required.
• Hexagonal wrench (nominal size 1.5)
• Spanner (nominal size 46)
(2) Turn off the power to the flowmeter.
(3) Using the spanner, loosen the hexagonal nuts at the
neck of remote sensor.
F0325.ai
(4) Using the hexagonal wrench, loosen the screw at the
neck.
(5) Rotate the terminal box in the desired direction.
NOTE
The terminal box can be rotated -140 degrees to
+180 degrees from the arrow mark indicating the flow
direction. Do not exceed these angle.
(6) Using the hexagonal wrench, retighten the screw at
the neck.
F0326.ai
+180 degrees
-140 degrees
(7) Using the spanner, retighten the hexagonal nut at the
neck, and then be sure to check that the terminal box
was firmly fixed to the sensor.
Installation
3