4
Introduction
Spurious Free Dynamic Range (SFDR)
Dynamic Range
System to Noise Ratio
Total Harmonic Distortion (THD)
SFDR (dBc)
1,2
Input Range (V)
Fs = 54 kS/s
Fs = 108 kS/s
Fs = 192 kS/s
±1V, ±10V
104
104
105
1. 1 kHz input tone and -1 dBFS input amplitude.
2. Measurement Includes harmonics.
Dynamic Range (dBFS)
1
Input Range (V)
Fs = 54 kS/s
Fs = 108 kS/s
Fs = 192 kS/s
±1V, ±10V
107
100
100
1. 1 kHz input tone and -60 dBFS input amplitude
SNR (dBc)
1
Input Range (V)
Fs = 54 kS/s
Fs = 108 kS/s
Fs = 192 kS/s
±1V, ±10V
104
99
98
1. 1 kHz input tone and -1 dBFS input amplitude
THD (dBc)
1
Input Range (V)
Fs = 54 kS/s
Fs = 108 kS/s
Fs = 192 kS/s
±1V
-106
-106
-107
±10V
-104
-104
-105
1. 1 kHz input tone and -1 dBFS input amplitude