GB 02329 - Edition 06 - September 12
1/2
B 310
Inlet and exhaust connections
Vacuum connections
Do not expose any part of the human body to vacuum.
The product is supplied with the inlet and exhaust sealed. Remove these
blanking plates when you are ready to connect the product on your vacuum
system.
As well as, don’t operate the product unless the inlet and exhaust are
connected to a vacuum and exhaust pumping line.
Risk of injury by cutting.
Contact with the pump rotor cell may cause cuts. Alternatively, protective
gloves may be worn when servicing the product.
Materials in direct contact with process gases
COMPONENTS IN THE PUMP
MATERIALS
From inlet to outlet
Stainless steel, aluminium alloy
O-ring
Fluorinated elastomers, FPM
Pump installation with inlet housing facing up:
- connect the pump to the customer’s handling device using the threaded holes
located at the rear of the pump (
A 400 of the Operating instructions).
After connecting the product to the pumping line, check for leaks along
the entire to ensure proper connections.
Pollution risk:
Unclean and contaminated component can increase the pumping down
time. Use only dry and clean pipe lines, and wear gloves to make the pump
connections.
Make sure all parts and chambers connected to the inlet, exhaust and
purge of the maglev pump can withstand a negative pressure of 100 kPa
below atmospheric pressure and that they are impervious to damage
from vacuum (seals, etc...).
For safety reasons, use accessories on the inlet and exhaust lines whose
materials and sealing properties are compatible with the gases being used.
adixen Vacuum Products - Operating instructions - ATH 2300 M/MT