Appendix C
Specifications
39
C
A P P E N D I X
Single Bio Amp
Input
Connection type:
Six-pin DIN/MS socket to fit 3-lead Bio Amp
cable (Tronomed D-1340)
Input configuration:
1 isolated differential channel with isolated
ground reference
Input impedance:
200 M
Ω
differential, 30 pF (no cable) or
500 pF (supplied Bio Amp cable and leads) to
isolated ground
Isolation: 4000
V
rms
(50 Hz for 1 minute)
Amplification ranges:
± 5 μV to ±100 mV full scale in 14 steps
(combined PowerLab and Bio Amp)
± 100 mV
± 50 mV
± 20 mV
± 10 mV
± 5 mV
± 2 mV
± 1 mV
± 500 μV
± 200 μV
± 100 μV
± 50 μV
± 20 μV
± 10 μV
± 5 μV
Gain accuracy:
± 1.5% all ranges