AD2817C Operation Manual
50
5.2.5 Data format
AD2817C outputs the measurement results using the ASCII via the GPIB bus. The Format 1
is used by
MeasDisplay
page,
BinNo. Disp
page and
Bin Count
page, while Format 2 is
used by
List Sweep
page. The two formats are described respectively as follows.
Figure 5-4 Data format 1
The <DATA A>, <DATA B>, and <BIN> formats are as follows.
l
<DATA A> and <DATA B> format:
The data output formats for <DATA A> (primary parameter's measurement data), and
<DATA B> (secondary parameter's measurement data) uses the 12 ASCII character fixed
length format as follows.
SN.NNNNNESNN
(S: +/-, N: 0 to 9, E: Exponent Sign)
l
<BIN> Format:
The <BIN No.> data shows the bin sorting results as follows.
Data
Sorting Results
1
BIN 1
2
BIN 2
3
BIN 3
4
AUX_BIN
5
OUT_OF_BINS
The <BIN No.> data is output with the measurement data only when the comparator
function is set to ON.