Carefully remove the delivery contents from the packaging. 

Put the load cross on the scale. Make sure that the holes in the centre of the load cross sit 
over the pin in the centre of the housing. Attach the load cross to the quick-release fastener. 
Place first the silicone mat and then the weighing plate on the load cross.  

Place the scale on a level and stable installation surface and make sure that all four feet are 
in contact with it. In addition, check the footing of the scale by means of the spirit level, the 
bubble of which should be in the centre of a circle. 

Put the weighing bowl on the scale. The four pins of the weighing bowl should grip into the 
appropriate scale openings. 

Insert the AC adapter into the socket located below the scale next to the power button. Connect 
the AC adapter to an AC power outlet (pay attention to the correct voltage). 

Using the scale 

Switching on the scale 

To turn on the scale, press and hold down the 

 button for about three seconds. 

The display will show “ADE” for two seconds and will then automatically switch to weighing 
mode. “0,000” is shown on the display.  

If this is not the case, press the 

 button. As a result, measurement inaccuracies will be 

compensated after turning on the scale and will be reset to zero before the actual weighing 
process. 
The display now shows “0.000” and the 

 figure indicates the zero position. 

 shows that 

the weight value is stable. The scale is now ready for use. 
If  the  display  shows  only  “0.00”,  please  check  the  selected  digit  interval  in  the  parameter 
settings of the unit of weight. 

Determining weight 

Make sure that the display shows “0.000”, 

 and 

 as described above. 

Put  the  object  to  be  weighed  on  the  weighing  surface.  The  weight  value  will  appear  on the 
display and 

 will show that the weight value is stable. 

Zeroing the scale 

If the weight is lower than 10% of the maximum capacity (here: 3 kg) and the display shows 

the stability symbol 

, press the 

 button to perform the zeroing function. 

The display will show “0.000”, 

 and 

Tare function 

If the weight value is greater than or equal to 10% of the maximum capacity (here: 3 kg) and 

the display shows 

, press the 

 button to perform the tare function. 

The display will show “0.000”, 

 and 

Summary of Contents for PWI30

Page 1: ...H Co KG Hammer Steindamm 27 29 22089 Hamburg Germany 49 0 40 432776 0 Fax 49 0 40 432776 10 E mail info ade germany de Portion scale PWI30 with induction charging PWI30 190920 Rev005 UM en IMPORTANT READ CAREFULLY BEFORE USE PLEASE KEEP FOR FUTURE REFERENCE ...

Page 2: ...harging 5 Assembling and levelling the scale 6 Using the scale 6 Switching on the scale 6 Determining weight 6 Zeroing the scale 6 Tare function 6 Setting the unit of weight 7 Switching off the scale 7 Error messages 7 Parameter settings 7 Backlight 8 Selecting the unit 8 Automatic power off in battery mode 9 Digit interval d 9 Zero offset 9 Cleaning 10 Safety information 10 Storage and transport ...

Page 3: ...PWI30 190920 Rev005 UM en 2 Overview display 1 Spirit level 2 LCD display 3 Buttons 4 Weighing plate 5 Housing 6 Foot 7 Induction charging base 8 Induction charger 9 Charging adapter ...

Page 4: ...rn on and off To toggle between the units of weight Display symbols EEEEEE Overload a beep sounds at the same time immediately remove the object from the weighing platform Indication that a fixed weight value has been stored which will be subtracted at each subsequent weighing process TARE symbol Indication that the scale is reset to zero Indication that the achieved weight value is stable Battery...

Page 5: ... power off as well as the display of different units can be set or disabled To achieve precise results please read the user guide carefully and follow the instructions contained therein Inspecting the contents of the box Portion scale Load cross silicone mat and weighing plate Quick release fastener for the load cross Induction charger User manual incl declaration of conformity Safety information ...

Page 6: ...ments symbols are opposite each other The charging function is available only if the distance between the transmission and reception elements is sufficiently small It is therefore necessary to ensure that the induction charger is properly positioned in the induction charging base LED display for induction charging Slide the induction charger into the induction charging base and connect the chargin...

Page 7: ... three seconds The display will show ADE for two seconds and will then automatically switch to weighing mode 0 000 is shown on the display If this is not the case press the button As a result measurement inaccuracies will be compensated after turning on the scale and will be reset to zero before the actual weighing process The display now shows 0 000 and the figure indicates the zero position show...

Page 8: ... back to the kg unit The selected unit is shown as on the display Switching off the scale To manually turn off the scale press and hold down the button for four seconds in weighing mode As soon as a beep sounds release the button and the scale will turn off Error messages Overload EEEEE display The scale overload 9d is signalled by the EEEEE display The overload must be eliminated immediately to p...

Page 9: ... unit Use this parameter setting to disable the units of weight the display of which is not required t9 Enable On or disable Off the display of kg as the unit of weight 9 Enable On or disable Off the display of g as the unit of weight lb Enable On or disable Off the display of lb as the unit of weight oZ Enable On or disable Off the display of oz as the unit of weight Press the button to save the ...

Page 10: ...e displayed digit interval It is recommended to leave this setting unchanged and use the button to switch immediately to the next parameter setting If it is necessary to adjust the division there are the following setting options d 1 1 g division d 2 2 g division d 5 5 g division d 10 10 g division d 20 20 g division Press the button to save the selected setting and switch automatically to the nex...

Page 11: ...ranes on the underside that prevent the formation of condensate inside the scale To ensure protection against mechanical damage the membranes are provided with two bleed screw caps that must also be cleaned if they are soiled During assembly make sure that the membranes and seals are inserted correctly Safety information Safe operation is no longer guaranteed If the AC adapter or the power cord sh...

Page 12: ...s etc are excluded This warranty does not cover external normal signs of wear and tear or damages caused by accident or misuse The warranty will not apply for devices that have been opened by unauthorised persons In addition the operation of the scale with damaged or missing membranes is not allowed and will void the warranty The statutory rights of the customer are not affected by this agreement ...

Page 13: ...afe Quick release fastener for disassembly of the weighing surface Working conditions Ambient temperature 10 C 50 C Relative humidity 1 80 Functions Weighing taring units of weight kg g lb oz User friendly 2 button operation Unit conversion 1 kg 1000 g 1 kg 2 206 lb 1 kg 35 27 oz Power supply 230 V AC mains and battery operation interface free Induction charging power adapter 5 V DC 1 A high perfo...

Page 14: ...tive 2014 30 EU regarding electromagnetic compatibility Low voltage directive 2014 35 EU RoHS Directive 2011 65 EC in the currently valid versions This declaration will lose its validity if any modification is made to the scale without our consent Hamburg January 2019 ADE Germany GmbH Co KG Hammer Steindamm 27 29 D 22089 Hamburg Manufacturer ADE Germany GmbH Co KG Hammer Steindamm 27 29 D 22089 Ha...

Page 15: ...PWI30 190920 Rev005 UM en 14 ...

Page 16: ...15 PWI30 190920 Rev005 UM en Distributor ...

Reviews: