ADE M70612 Operating Manual Download Page 5

060100-Rev000-M70612  

 
 

Tare Function 

 

Use the 

TARE key 

when the patient is weighed with any other equipment (such as carrying equipment, 

belts, suspension yoke etc.). Weight the equipment first. After the weight is indicated press 

Tare 

key to 

adjust the scale back to 

zero

. The weight of the stored value will be deducted for all future weighing 

procedures unless the tare function is used another time.  
 
The weight displayed in the patient’s net weight. 

 
 

 

GROSS/NET function 

 

Using the 

GROSS/NET key 

you can jump between gross and net subject to a tare value has been 

stored. 
 
 

Zero setting 

 
Pressing the 

Zero key 

you may

 

eliminate minor weighing irregularities which may occur after starting 

the scale and the scale will be set to zero. 

 
 
 

Printing function 

 

In case a printer has been connected and the printing menu has been activated  the<

PRINT

> function is 

sending the weight result to the printer. In case no printer has been connected the function is 
interrrupted after 10 seconds. 

 
 
 

„Lifeweight“ - function 

 

The

 

f- key

 is manually activated and deactivated dependant from the requirement to weigh extremely 

restless patients and getting a reliable result. In case the Lifeweight function is activated the scale  
identifies all major movements caused by the patient during weighing in either direction and is 
automatically cutting these irregularities off. Different automatical measurements done during weighing 
the final weight of the patient displayed is most accurate. 

 

 
Display symbols 

 
 

NET 

                     

Scale is under tare 

(Motion) 

 

Scale is in motion 

 

S

 

(above limit)              The limit has been exceeded 

 

T

 

(below limit)              Off limit 

 

 

(Battery) 

 

Batteries low – insert fresh ones 

 