valuating Body Mass Index 

 

 

 

 

 

 

 

Print out a record 

The scales must be at standstill, e.g. standstill status display 

 (14) appears at the screen. 

Press the ENTER key (13) to send the data to the serial interface. 
Note:  Printing  is not possible  if the  net  weight is 0. The print function  only  produces one  print-out.  A reset is 
required to print out another copy.  
 

Care, servicing and recalibration 

Cleaning 

Remove the mains plug before cleaning the  unit. Do  not use any aggressive cleaning materials for cleaning; 
only use a damp cloth or conventional disinfectant. Ensure that no liquids or any fluid gets into the scale.

 

Error messages 

Error mes-
sage 

Description 

Problem solution 

  
 

Overload display 

Weight exceeds max. capacity of the 
scales. 

 

Overflow indicator (negative) 

Negative weight indicated – weight too 
low to be displayed (<=99999). 

 

Overflow indicator (positive) 

Positive weight indicated – weight too 
high to be displayed (>=999999). 

AD LOW 

A/D above negative range 

Check if the scales are correctly as-
sembled or if they are damaged. 

AD HI 

A/D above positive range 

Check if the scales are correctly as-
sembled or if they are damaged. 

EE SUM 

Parameter or calibration error 

Recalibrate – please contact your sup-
plier or service partner: 

EE WR 

EEPROM indicates an error 

Please contact your supplier or service 
partner. 

PM SUM 

Internal program indicates a check error 

Please contact your supplier or service 
partner. 

HOFSET 

Switch on – above zero 

Please remove the weight from the 
scales and restart. 

LOFSET 

Switch on – below zero 

Check scales for damage and inform 
your service partner: 

UOFSET 

Irregular 

Do not touch the scales after they have 
been switched on. 

 

BMI evaluation according to WHO 

BMI   

<  16 

Severely underweight 

  16 

17 

Moderately underweight 

  17 

18.5  Slightly underweight 

BMI  18.5 

25 

Normal weight 

BMI  25 

30 

Overweight 

  30 

35 

Obese Class I 

  35 

40 

Obese Class II 

   

>  40 

Obese Class III 

 

Summary of Contents for M500020 Series

Page 1: ...M500020 210322 Rev009 UM e Operating Manual Electronic Wheelchair and Multipurpose Scales Series M500020...

Page 2: ...devices 7 Housing and connections 7 Safety 8 Safety instructions 8 Settings 8 Setting date and time 9 Operation 9 Modes of operation 9 Weighing 9 Tare function 10 Hold function 10 Determining the Body...

Page 3: ...ative or visit our web page at www ade hamburg de This operation instruction applies to the following models of electronic scales of the M500020 series with the different adaptations see table 1 In or...

Page 4: ...tor display of your ADE scale is able to be mounted on walls on sockets handrails or on stand on devices for the table depending on the model you have selected Please insure that your scale is being p...

Page 5: ...power unit The imprinted voltage value at the mains adapter must match the local mains voltage Connect the mains cable to the indicator and to a socket The wall socket must be located near to the uni...

Page 6: ...geable batteries into the domestic waste Ensure that the valid legal and municipal regulations of your country are being followed Never dispose of batteries in fire as they can leak or explode Quality...

Page 7: ...right Key 4 menu direction to the left Key 8 menu direction down Key 2 menu direction up For better understanding see Pic 1 1 the blue arrows Used to navigate through menus to select numerical values...

Page 8: ...ld match the mains voltage and power type at the place of use Make sure that the mains cable does not pose any potential trip hazard Do not operate the unit in hazardous or unstable conditions Do not...

Page 9: ...e Date time and consecutive numbers for print commands are being set in this mode To start setting mode hold the GROSS NET key pressed until TIME appears on the display Configuration mode set up mode...

Page 10: ...o change the already displayed value within five sec onds the unit will turn back to weighing mode and displays the last stored value as negative figure as above Enter the on off zero button 8 to leav...

Page 11: ...99 Overflow indicator positive Positive weight indicated weight too high to be displayed 999999 AD LOW A D above negative range Check if the scales are correctly as sembled or if they are damaged AD H...

Page 12: ...ar damages caused by incorrect or negligent handling overuse or unsuitable or improper storage The cost of transport will be borne by the customer should the equipment be located anywhere other then t...

Page 13: ...Full duplex RS 232 38400 19200 9600 4800 2400 1200 bps Printer Port Output only RS 232 or active 20 mA current loop 9600 4800 2400 1200 bps Both Ports 7 or 8 data bits even odd or no parity Operator...

Page 14: ...formity by the manufacturer This product has been manufactured in accordance with the harmonised European standards They con form to the regulations of the EC directives listed below NAWI Directive 20...

Page 15: ...M500020 210322 Rev009 UM e 15...

Page 16: ...16 M500020 210322 Rev009 UM e Dealer...

Reviews: