Heating additional compules:

3.  Do not turn the unit off.

4.  Remove the first compule.

5.

  Install a new compule and immediately turn the unit off and then 

back on.

6.  Allow the compule to heat for one minute. At one minute the temperature 

status LED will turn from blinking AMBER to blinking GREEN and the 

composite is warm and ready to express.

7.  Additional compules can be heated by repeating steps 3 through 6.

Optional:  Save time by pre-heating additional compules using the 

AdDent Calset Composite Warmer.

1.  Set the Calset Composite Warmer on the high temperature setting 155°F 

(68°C) and allow it to reach operating temperature (20 minutes for small 

trays, 30 minutes for large trays).

2.  Place the desired number of compules in the Calset Tray and allow 

sufficient heating time, approximately three minutes.

3.  

Do not turn the Compex unit off.

4.  Remove the first compule.

5.  Install a pre-heated compule in the Compex. T

he composite is ready to 

express immediately.

6.  Additional compules can be heated by repeating steps 3 through 5.

• 

Note

 - We suggest you try shorter and longer heating cycles to determine 

the optimum heating time for the composite you are using prior to clinical 

application.

• Once the heating cycle has been activated, pressing the on/off button again 

will shut off the unit and switch it back into sleep mode.

• The auto shut off feature takes place after 10 minutes of ON time to 

conserve battery power.

• Tip:  Avoid touching the compule ejector to the patient while dispensing 

material to avoid accidently ejecting the compule.

• Tip:  To conserve battery power, turn off dispenser after procedure is 

complete.

 

     Caution:  To avoid damage to the unit, always remove compule after  

 

     procedure is complete.  Do not store the Compex HD unit with    

compule installed in the compule holder.

COMPULE HEAT UP TIME

• The Compex will heat the composite from 72°F (22°C), room temperature, 

to 155°F (68°C) in approximately 2 minutes.

REMOVING THE COMPULE

 

             Caution. Some compules can forcefully eject during the removal  

             process. Be sure to hold your hand below the compule to catch it 

when it comes out.

• To remove the compule, press down on the raised portion of the compule 

ejector. This feature allows the rear of most compules to tilt down for easy 

removal.

CLEANING/DISINFECTING/STERILIZING

a)  Cleaning the silicone sleeve:

CHARGING THE UNIT

Note: The Compex HD unit should be fully charged prior to first time use.

A wall plug-in USB power supply is provided to charge the unit. The power 

supply operates from wall Voltages from 100 to 240 Volts, AC.  Connect the 

USB cord to the power supply and plug the power supply into a suitable wall 

outlet. Plug the opposite end of the USB cord into the connector located on 

the bottom of the Compex HD handle.

 

       

Do not attempt to use the Compex HD when the unit is   

 

 

      charging. The unit cannot be activated while the power supply 

is connected to the unit.

BATTERY CHARGE INDICATOR

             Charging the Compex HD:  While charging the unit, the right side        

             LED battery charge indicator blinks red.  When the unit is fully 

charged, the LED turns solid green.

Operating the Compex HD:

  While using the unit, the right side LED battery 

charge indicator operates as follows:

GREEN - Battery is charged, normal operation.

AMBER - Battery charge at approximately 20%, unit has 10 minutes of 

useful charge remaining.

RED - Battery is at the end of useful charge and requires recharging.

 

 

TEMPERATURE STATUS INDICATOR

 

      The LED on the left side indicates the temperature status.  For the    

 

       first minute of the heating cycle the temperature status LED blinks    

             AMBER.  At one minute the LED starts blinking GREEN.  At two 

minutes the LED turns solid GREEN.  The LED stays solid GREEN for the 

remaining eight minutes of the heating cycle.

INSTALLING THE COMPULE (PLT)

1.  Place the compule into the opening in the barrel ensuring that the 

compule flange is positioned as far back as possible.

2.  Press the compule into the compule holder with firm even pressure.  The 

compule will snap into place.  

3.  Ensure that the compule body is fully seated against the compule holder 

with the compule flange as far back as possible.

OPERATION – HEATING AND DISPENSING COMPULES

•  

The heating cycle warms the composite to 155°F (68°C):

  Pressing the 

on/off button once activates the heating cycle. For the first minute of the 

heating cycle the temperature status LED blinks AMBER.  At one minute the 

LED starts blinking GREEN.  At two minutes the LED turns solid GREEN.  

The LED stays solid GREEN for the remaining eight minutes of the heating 

cycle.

Heating one compule:

1.  Install a compule and press the on/off button once.

2.  Allow the compule to heat for two minutes.  At two minutes the 

temperature status LED will turn solid GREEN and the composite is warm 

and ready to express.

!

• Immerse/thoroughly drench parts in a surface decontaminate cleaner 

(soap and water cleaning solution).

• After 10 minutes, scrub with an instrument cleaning brush until visibly free 

from debris.

•  Rinse all parts with warm water.

b)  Sterilizing the silicone sleeve:

• Sterilize in a steam autoclave using distilled water at a temperature of 

270°F (132°C) for 20 minutes.

• Do not add chemical solutions to the autoclave.

• Do not use a chemical autoclave.

c)  Disinfecting: The Compex HD housing can be disinfected following these 

guidelines:

• Wipe all surfaces with a disinfectant comprising 72% - 55% alcohol and 

.33% -.5% quaternary ammonium and follow the manufacturer’s directions. 

Some popular disinfectant wipes in this category are:

1 - Monarch Surface Disinfectant Wipes (Air Techniques)

2 - Opticide 3 Wipes (Biotrol)

3 - Super Sani-Cloth Wipes (PDI)

• Wipe, do not spray, solution onto the Compex HD housing.  Prevent liquids 

from entering openings on unit.

        Do not autoclave the Compex HD unit. Doing so will damage the  

         unit and invalidate the warranty.

In order to protect this device and help prevent cross contamination we have 

developed a disposable barrier bag and an autoclavable silicone sleeve.

TROUBLESHOOTING GUIDE

!

!

!

          Problem                  

1. Battery not charging 

2. Trigger sticking after 

expressing composite 

material. 

3. Compule not heating 

4. Compule not heating 

- left side LED heat 

indicator turns RED for 

five seconds then turns 

off.

          Cause 

                

a. USB Charging cable 

connection 

b. No power to the USB 

Power Supply.  

c. Battery at end of life

Composite 

accumulated on 

plunger tip

a. Composite material 

accumulated on heating 

element. 

b. Battery discharged 

(battery indicator red or 

unit will not turn on) 

c. Unit has turned off 

after 10 minutes. 

Faulty temperature 

sensor

            Solution                  

a. Ensure USB charging 

cable is securely 

connected to USB 

power supply and to the 

Compex HD unit.  

b. Ensure there is power 

at the USB power supply. 

c. Replace the battery 

Squeeze trigger to 

expose plunger tip and 

use soft cloth or Q-tip 

to remove accumulated 

composite material. 

a. Use soft cloth or Q-tip 

to remove accumulated 

composite material. 

b.  Charge battery 

c. Restart heating cycle

Return unit for repair.