Assembly
Your ADC DIAGNOSTIX
TM
e-sphyg
TM
digital aneroid sphygmomanometer consists of a
digital manometer (gauge), complete inflation system (ADCUFF
TM
calibrated nylon cuff,
latex inflation bladder (latex-free bladder optional), squeeze bulb, and the ADFLOW
TM
valve), a zippered carrying case, and operating instructions.
Parts and Features
• Extra large LCD is easier to read
• Virtually indestructible
• Backlight display for viewing in any light condition
• Splash resistant housing for use in harsh environments
• Manometer works with virtually any cuff
Gauge Hang
Tab
Arm Cuff
Gauge
Backlight
Button
On / Off
Zeroing Button
Artery Mark
Air Tube and
Connector
Air Release Valve &
Inflation Bulb
Splash
Resistant
Housing
Extra
Large LCD
Display
Error Message:
Description: Solution:
ERROR DISPLAYS and TROUBLESHOOTING
Err
Heartbeat rate
detection error.
The cuff was deflated
too rapidly.
Keep the deflation
rate at about 2-3
mmHg per second.
EE
Press the “ ” button
to switch the unit off,
or the unit will
automatically switch
off after
60 seconds.
E3
Data Error.
Remove and reinsert
the batteries. If this
error keeps occurring,
return the device to
your local distributor
or service center.
01
The cuff was inflated to
a pressure over 300
mmHg.
Open the valve
immediately to
rapidly exhaust the
air in the cuff.
A Pressure of over
15mmHg remaining
for more than
3 minutes is detected.