background image

12

5

Care and Maintenance

Storage:

After measurement, wrap cuff around gauge and bulb and store in zippered carrying case.

Cleaning: Digital Manometer

: Your ADC

®

DIAGNOSTIX

TM

brand e-sphyg

TM

gauge requires

minimal care and maintenance.

The manometer may be cleaned with a slightly dampened soft cloth, but should not be
dismantled under any circumstances. 

Cuff:

Sponge with a damp cloth or alcohol pad.  If necessary, it can be washed in cool

water with mild soap or detergent-disinfectant. Remember to remove bladder from cuff
prior to washing. After washing, it should be rinsed and allowed to air dry. 

NEVER IRON.

Be sure to remove the bladder and all components before cleaning the cuff.

Bladder and Bulb:

Remove bladder from cuff and disconnect bulb from tubing and valves.

Wipe with a damp cloth. To help preserve it from cracking or discoloring, wipe with a soft
cloth moistened with ethanol.

Valves:

Clean the surface with a dry fabric such as soft cotton material. Remove dust

from filter screen.

Carry Case:

Wipe with a damp cloth and dry thoroughly with a dry cloth. If

necessary, it can be washed with cool water and mild soap. Do not allow to air dry.

Batteries:

If the e-sphyg

TM

will not be used for an extended period of time, remove batteries.

Disinfection and Sterilization

Disinfecting:

Glutaraldehyde-type liquid disinfectant may only be used on the cuff.  Follow

instructions for use provided with the glutaraldehyde product. Prolonged use of this disin-
fectant may cause discoloration. DO NOT use glutaraldehyde-type liquid disinfectant on
the digital gauge, bulb, valves, or bladder.

Sterilizing:

The blood pressure cuff may be gas sterilized. 

DO NOT

use steam or heat to

sterilize the cuff, bulb, valve, or bladder. 

DO NOT

attempt to sterilize the digital gauge.

WARNINGS

A warning statement in this manual identifies a condition or practice which, if not corrected
or discontinued immediately could lead to patient injury, illness, or death.

NOTE:

Systolic Assist and pulse measurement functions are NOT motion tolerant. 

(Pressure readings ARE). In a dynamic environment, systolic assist and pulse rate 
display may not operate correctly.

Caution:

Federal law restricts this device to sale by or on the order of a physician 

or licensed healthcare practitioner.

Caution:

Leaky batteries can damage the unit.

Caution:

Remove the batteries when the unit will not be used for an extended period 

of time.

Caution:

Your digital aneroid is a medical device designed for use by trained  

medical professionals. Keep out of the reach of children.

Caution:

Your digital aneroid is not field serviceable.

Caution:

You should not use any tool to open the device nor should you attempt to

adjust anything inside the device.

Caution:

To stop the operation of your digital aneroid at any time, open the air 

release valve as you would on any mechanical blood pressure instrument, and the 
air in the cuff will be rapidly exhausted.

Caution:

Your digital aneroid is designed for normal blood pressure measurement 

at the upper arm or thigh on healthy skin. Proper use of this device is essential for 
accurate measurement.

Caution:

Do not inflate your digital aneroid unit to more than 300 mmHg.

Warning:

If luer lock connectors are used in the construction of tubing, there is a 

possibility that they might be inadvertently connected to intravascular fluid systems, 
allowing air to be pumped into a blood vessel. Immediately consult a physician if this
occurs.

Warning:

Do not allow a blood pressure cuff to remain on patient for more than 10 

minutes when inflated above 10 mmHg. This may cause patient distress, disturb 
blood circulation, and contribute to the injury of peripheral nerves.

Warning:

Safety and effectiveness with neonate cuff sizes 1 through 5 is not established.

Summary of Contents for E-Sphyg 7002

Page 1: ...m email info adctoday com 2007 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the expressed written permission of ADC ADC ass...

Page 2: ...he finest sphygmomanometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to provide you with many years of depend able service Please read the followi...

Page 3: ...normal use and service as follows 1 Warranty service extends to the original retail purchaser only and commences with the date of delivery 2 The entire sphygmomanometer is warranted for one year 3 The...

Page 4: ...and accuracy of your new digital e sphygTM gauge Using state of the art technology your e sphygTM will provide digital accuracy with tank like reliability Intended Use ADC aneroid digital aneroid and...

Page 5: ...A warning statement in this manual identifies a condition or practice which if not corrected or discontinued immediately could lead to patient injury illness or death NOTE Systolic Assist and pulse m...

Page 6: ...utton On Off Zeroing Button Artery Mark Air Tube and Connector Air Release Valve Inflation Bulb Splash Resistant Housing Extra Large LCD Display Error Message Description Solution ERROR DISPLAYS and T...

Page 7: ...nician is qualified to perform this calibration service Do not attempt to recalibrate the gauge yourself In addition to the visual inspection indicated above we also recommend you perform the followin...

Page 8: ...the backlight off at any time press the button on the left side You will need to deactivate the backlight each time you power on the instrument Assembling your Sphygmomanometer Your inflation system i...