Start inflation
– Press the Start/Stop Button to begin inflation. When maximum
inflation pressure is reached, the e-sphyg 3 will automatically begin linear
deflation at a rate of 3 mmHg/sec.
Re-inflate
– Push and Hold the Memory Button during deflation to re-inflate
for as long as the button is held (up to a max of 299 mmHg). Release the button
to continue deflation.
Exceeding 299 mmHg will result in an immediate release of cuff pressure and a
'
HI
' error message.
When the cuff pressure reaches 20 mmHg during the deflation cycle, the
remaining pressure is vented and the e-sphyg 3 goes into standby mode.
Push the Start/Stop Button at any time to start fast deflation and set the
e-sphyg 3 to standby.
Take note
of the systolic and diastolic values in the same manner as
performed with sphygmomanometer measurements.
Set to standby
– The device can be set to standby while a reading is
displayed by pressing the Start/Stop Button. The device will automatically
switch to standby if there is no operation for one minute.