Assembling your Sphygmomanometer

Your inflation system is pre-assembled and ready for use. You will need to simply

connect your digital gauge to the complete inflation system by inserting the twist

lock connector (pre-assembled on a bladder tube) into the air port at the bottom of

the digital gauge, twist to lock.

Measurement Procedure

(Auscultatory Method)

    

1. Patient Position

: The patient should sit or lie comfortably. The arm should be

fully  supported on a flat surface at heart level. (If the arm’s position varies, or is

not level with the heart, measurement values obtained will not be consistent with

the patient’s true blood pressure.) 

    2. Apply the cuff:

To attach the digital gauge, plug the connector into the gauge

and twist clockwise to ensure a secure connection. The e-sphyg

TM

gauge can be ei-

ther clipped on the gauge hang tab, or held in your hand during use. ADCUFF

TM

nylon cuffs are specially designed to promote the accurate determination of blood

pressure. Index and range markings ensure use of the correct cuff size. The artery

mark indicates proper cuff positioning.

    Place the cuff over the bare upper arm with the artery mark positioned directly

over the brachial artery. The bottom edge of the cuff should be positioned approxi-

mately one inch above the antecubital fold. Wrap the end of the cuff not containing

the bladder around the arm snugly and smoothly and engage adhesive strips. To

verify a correct fit, check that the Index Line falls between the two Range Lines.

    3. Activate the digital aneroid and inflate the cuff:

Press the “       ” On/Off but-

ton to activate your digital gauge. After zeroing, your digital gauge is ready to meas-

ure. Close the ADFLOW

TM

valve by turning thumbscrew clockwise. Palpate the

radial artery while inflating the cuff. Be sure to inflate cuff quickly by squeezing

bulb rapidly. Inflate cuff 20-30 mmHg above the point at which the radial pulse disap-

pears.

Summary of Contents for Diagnostix E-sphyg 7002

Page 1: ...1 e sphygTM User s Manual DigitalAneroid...

Page 2: ...ormance You ll quickly appreciate the results for you now own one of the finest sphygmomanometers that money can buy With proper care and maintenance your ADC blood pressure instrument is sure to prov...

Page 3: ...d Maintenance 12 Disinfection and Sterilization 12 Specifications 13 Standards 13 Limited Warranty 14 Product Registration 15 This manual is intended to assist the user in the safe and efficient opera...

Page 4: ...digital e sphygTM aneroid also has several features not available in a more traditional mechanical gauge The systolic assist function will give you a visual alert when you are ap proaching the patient...

Page 5: ...igned for normal blood pressure measurement at the upper arm or thigh on healthy skin Proper use of this device is essential for accurate measurement Caution Do not inflate your digital aneroid unit t...

Page 6: ...ometer consists of a digital manometer gauge complete inflation system ADCUFFTM calibrated nylon cuff latex free inflation bladder squeeze bulb and the ADFLOWTM valve a zippered carrying case and oper...

Page 7: ...acing in the correct direction 4 Replace the battery cover and tighten the screw 5 Rotate the clip back Note You should replace the batteries if The weak battery mark appears in the display Nothing ap...

Page 8: ...our hand during use ADCUFFTM nylon cuffs are specially designed to promote the accurate determination of blood pressure Index and range markings ensure use of the correct cuff size The artery mark ind...

Page 9: ...display this information on the LCD screen once the cuff has been deflated Note Systolic assist and pulse display features are NOT motion tolerant Exces sive movement may interfere with accurate displ...

Page 10: ...per second and stop to check at the following test points 299mmHg 250 mmHg 150 mmHg 100 mmHg and 50 mmHg Your gauge should be within 3 mmHg of your reference stan dard ADC recommends using the most se...

Page 11: ...to switch the unit off or the unit will automatically switch off after 60 seconds E3 Data Error Remove and reinsert the batteries If this error keeps occurring return the device to your local distribu...

Page 12: ...a damp cloth To help preserve it from cracking or discoloring wipe with a soft cloth moistened with ethanol Valves Clean the surface with a dry fabric such as soft cotton material Remove dust from fil...

Page 13: ...ure 32 115 F 0 46 C Humidity 85 Relative Humidity max STORAGE ENVIRONMENT Temperature 4 131 F 20 55 C Humidity 90 Relative Humidity max non condensing Dimensions 3 L X 23 8 W X 11 8 H Weight 3 14 oz G...

Page 14: ...by abuse mis use accident or negligence Incidental special or consequential damages Some states do not allow the exclusion or limitation of incidental special or consequential damages so this limitati...

Page 15: ...00 232 2670 fax 631 273 9659 www adctoday com email info adctoday com 2015 ADC All rights reserved No one is permitted to reproduce or duplicate in any form this manual or any part thereof without the...

Reviews: