5

  
 

Medical Disclaimer

 

 

The ADC@Home app is not intended for use in the diagnosis, cure, treatment, 
or prevention of disease. It is not intended for emergency use or real-time mon-
itoring, nor as a substitute for medical care by a healthcare provider. ADC does 
not make, advise, or influence treatment decisions. Additional information and 
limitations are available on our website at 

www.adctoday.com/ADCatHome

.

 

Symbol   Definition

 

                

                 Important Warning/Caution 

 

                 Relative humidity limits 

 

                  Lot code 

 

                 Not made with natural rubber latex 

 

                 Phthalate free 

 

                 Keep dry 

 

                 Type BF applied part 

 

                 Manufacturer’s information 

 

                 Recycle the product separately 
                 from other disposables  

 

                 UDI number

 

 

                 Medical device

Symbol   Definition

 

                

                 Temperature limits 

 

                 Product identifier 

 

                 Limb circumference 
                (Minimum/Maximum) 

               

Non-ionizing electromagnetic 

                radiation 

               

Authorized European 

                  representative’s information 

               

Meets essential requirements of 

                European Medical Device Directive 
                93/42/EEC 

               

Protected against solid foreign 

                  particles with a diameter of more 
                  than 12.5 mm, no protection 
                  against water

IP20

 Symbol Definitions

 

The following symbols are associated with your Advantage Connect.

0044

Summary of Contents for advantage connect 6024N

Page 1: ...cal instrument must be used according to instructions to ensure accurate readings Questions Call ADC at 1 800 232 2670 Instructions for Use Models6024Nand6025N advantageconnect Featuring ADC HomeTM Connectivity ...

Page 2: ...er Model 6025N Only 11 Set the Time and Date 12 Connect the Cuff 12 Select the Measurement Mode and User 13 Using Your Monitor with a Smartphone or Tablet 14 Install the App 14 Set Up Your Profile 14 Pair Your Device 15 Preparing for a Measurement 16 17 6024N Fitting the D Bar Cuff 16 17 6025N Fitting the Rigid Cuff 17 Taking a Measurement 18 19 Stopping a Measurement 19 Using the Memory 20 21 Vie...

Page 3: ...ffices throughout the world where accuracy and dependability are critical ADC professional diagnostic products are the in struments of choice Now you too can enjoy the benefits of ADC engineering and quality in the home This instrument was designed to simplify the measurement of blood pressure and pulse rate at home and deliver consistent dependable results that can be stored and managed on your s...

Page 4: ...th a user friendly design Before getting started please read this instruction manual carefully and then keep it in a safe place Additional documentation is available at www adctoday com ADCatHome Please contact your doctor if you have further questions about blood pressure and its measurement Contraindications This device is contraindicated for any person who is connected to a wearable or implanta...

Page 5: ...y limits Lot code Not made with natural rubber latex Phthalate free Keep dry Type BF applied part Manufacturer s information Recycle the product separately from other disposables UDI number Medical device Symbol Definition Temperature limits Product identifier Limb circumference Minimum Maximum Non ionizing electromagnetic radiation Authorized European representative s information Meets essential ...

Page 6: ... that is transmitted with the app must be eval uated by your physician at a later time WARNING The device is not suitable for use on pregnant women users with implanted electric devices users with preeclampsia premature ventricular beats atrial fibrillation or peripheral arterial disease users undergoing intravas cular therapy or arteriovenous shunt or people who received a mastectomy Consult with...

Page 7: ...it is not suitable for continuous monitoring during medical emergencies or operations WARNING Taking blood pressure measurements too frequently could disrupt blood circulation and cause injuries WARNING Do not kink the connection tube during use The cuff pressure might continuously increase which could prevent blood flow and result in injury WARNING Never apply the cuff over damaged skin This can ...

Page 8: ...manually enter blood pressure pulse rate and weight readings on your app confirm that the numbers are accurate before saving them CAUTION To ensure the privacy of your medical information and to protect your self against cyber threats review and adjust the security settings on your phone CAUTION To ensure the privacy of your medical information only send reports with your medical information to pe...

Page 9: ...e range D bar nylon cuff 850 6022N Fits arm circumference from 22 cm to 42 cm 8 7 in to 16 5 in 6025N Self adjusting wide range rigid cuff 850 6023N Fits arm circumference from 22 cm to 46 cm 8 7 in to 18 1 in x3 4 AAA Batteries Instructions for Use Blood Pressure Monitor Model 6025N Also Includes Battery Compartment A C Power Adapter Carrying Case What s in the Box ...

Page 10: ...ment Err 3 Cuff Pressure Date Time Indicator AM PM Indicator Date Time Systolic Value x3 Diastolic Value Pulse Rate Irregular Heartbeat Detection Bluetooth Pairing Reading Value against Traffic Light Cuff Signal Indicator Battery Strength MAM Indicator Display Icons ...

Page 11: ...M Low Battery Warning Expired Battery Warning Use AAA long life or alkaline 1 5V batteries Do not use rechargeable batteries Remove batteries if monitor will not be used for long periods i i The Advantage Connect Model Plus Model 6025N only can be used with batteries or the included AC adapter No power is taken from the batteries while the AC adapter is connected to the device Use only the AC adap...

Page 12: ...shing M SET Press until you reach the current year Press to confirm Repeat Steps 1 and 2 to enter current month date hour and mi nutes Note the AM and PM icons when setting the hour Year Unit will accept the year and continue to month setting 1 2 3 Connect the Cuff Insert the cuff connector into the port on the left side of the unit 1 ...

Page 13: ...x3 1 x3 Press and hold for 5 seconds SET Press to confirm Average Mode SET Press to confirm user number 2 1 Flashes M Press to turn Average Mode on or off For more information on MAM Advantage Average Mode see page 22 Or press to go directly to user number settings Step 4 SET 1 x3 User setting flashes next M Press to toggle user number i 1 2 3 4 5 Repeat these steps at any time to change the user ...

Page 14: ...iew graph annotate and share reading with your physician or caregiver If you plan to use the ADC Home app complete this setup before taking a measurement For additional instructions on using the app see www adc com ADCatHome Install The App Open the app on your device then tap Create Account Fill in required fields First Name Last Name Email Password and tap Create Account i 1 1 2 Find and install...

Page 15: ...ablet any time you take a reading and have the app open If the monitor is not able to pair within two minutes Bluetooth will auto matically turn off Press the SET button on the monitor to try again Open the app and check that Bluetooth is on and discoverable Pair Your Device On the blood pressure monitor 1 2 3 i Troubleshooting tips are available at www adctoday com ADCatHome i i If prompted in th...

Page 16: ...n a bare arm if possible Always measure on the same arm normally left Try to take measurements around the same time each day o 3 o 3 o 3 For additional tips and information on obtaining accurate readings see 25 Loop cuff through ring 6024N Fitting the D Bar Cuff Position cuff over left arm Tighten cuff 3 4 to 1 above elbow Red Artery Mark over artery Tube toward lower arm Pull end and close hook a...

Page 17: ...it Test fit by sliding two fingers between arm and cuff White triangle falls between range lines 4 1 2 3 If the cuff provided with your unit is too big or small please call 1 800 232 2670 Only use approved ADC cuffs i If using the ADC Home app open the app on your phone or tablet be fore continuing to the measurement i If the cuff provided with your unit is too big or small please call 1 800 232 2...

Page 18: ...inflation stops and cuff deflates x3 if MAM Average Mode is on countdown to the second or third reading will appear and the next reading will start 1 If using the ADC Home app open the app on your phone or tablet be fore starting the measurement For information on setting up the app see pages 14 15 i You ll hear a slight pause when inflation nears 30 mmHg as the inflation process calibrates for op...

Page 19: ...to pair within two minutes Bluetooth will disconnect and the device will turn off Stopping a Measurement You can stop a measurement at any time Press and release Unit will turn off and cuff will deflate If you experience any discomfort such as pain in the arm stop the measurement immediately and remove the cuff i i 2 1 When measurement has concluded the systolic pressure diastolic pressure and pul...

Page 20: ...re viewing the measurements for the correct user Once the maximum memory capacity of 99 is reached the next new reading will overwrite the oldest value currently in memory This only affects readings that are stored on the monitor itself data within the ADC Home app is not impacted 2 1 M x3 Average of all stored values for that user appears first Press and release M Then the measurement appears Ind...

Page 21: ...to the data in the future Use the ADC Home app or keep a written record if needed 2 1 Confirm that the monitor is set for the correct user For more informationonchangingtheuser seepage13 Press and hold for 5 seconds M Press to confirm SET i 1 2 3 Press and release to view the next stored reading M 3 4 Clearing the memory only affects readings that are stored on the monitor itself if you are using ...

Page 22: ... from single measurements x3 For more information on turning Average Mode on and off see page 13 If an error occurs during any of the measurements that reading will be discarded and an additional measurement is taken automatically Do not remove the cuff between measurements There is a break of 15 seconds between the measurements An on screen countdown indicates the remaining time before the second...

Page 23: ... a regular basis e g several times a week with measurements taken daily we advise you to speak with your doctor Information for the doctor on frequent appearance of the Irregular Heartbeat Detector This device is an oscillometric blood pressure monitor that also analyzes pulse irregularity during measurement The device is clinically tested The arrhythmia symbol is displayed after the measurement I...

Page 24: ... mm Hg and or Diastolic mm Hg Category upper number lower number HYPERTENSIVE CRISIS Higher Than 180 and or Higher Than 120 Consult your doctor immediately HIGH BLOOD PRESSURE 140 or Higher or 90 or Higher HYPERTENSION Stage 2 HIGH BLOOD PRESSURE 130 139 or 80 89 HYPERTENSION Stage 1 ELEVATED 120 129 and Less than 80 NORMAL Less Than 120 and Less than 80 Only a physician is qualified to interpret ...

Page 25: ...results in a measurement error of 10 mmHg Cuffs that are too narrow or too short result in false measurement values Selecting the correct cuff is extremely important Cuff size is dependent upon the circumference of the arm measured in the center Each model has its own cuff 6024N comes with cuff 850 6022N which fits arm circumference from 22 cm to 42 cm 8 7 in to 16 5 in 6025N comes with cuff 850 6...

Page 26: ...ch can lead to false results Consecutive measurements in average mode are repeated with a 15 second pause between readings If you want to take a second set of average mode readings be sure to wait at least one minute before starting the sequence again To learn more about blood pressure see our informational document About Blood Pressure at www adctoday com ADCatHome i i i i ...

Page 27: ...ignal too weak The pulse signals on the cuff are too weak Reposition the cuff and repeat the measurement Err 2 Error Signal During the measurement error signals were detected by the cuff caused for instance by movement or muscle tension Repeat measurement keeping arm still Err 3 Abnormal cuff An adequate pressure cannot be generated in the pressure cuff A leak may have occurred Check that the cuff...

Page 28: ...s with the monitor contact the dealer where you bought the device or ADC Never attempt to repair the in strument yourself Unauthorized opening invalidates the warranty Other possible errors and their solutions If problems occur when using the device the following points should be checked Malfunction Remedy 1 Check battery installation polarity 2 Remove the batteries and if the display is unusual t...

Page 29: ...refully with a damp cloth and soapsuds Do not drop the monitor Avoid stress when handling the cuff and tubing Twisting or buckling can damage the sensitive air tight bubble Keep away from sharp edges Do not clean with gas thinners or similar solvents Do not submerge any parts in water Do not expose to extreme temperatures humidity dust or direct sunlight Never open the monitor This invalidates the...

Page 30: ... to Korotkoff method Phase I systolic Phase V diastolic Measurement range Sys 60 225 mmHg Dia 40 200 mmHg Pulse 40 199 beats per minute Memory 99 measurements for each user Cuff pressure display range 0 299 mmHg Measuring resolution 1 mmHg Static accuracy Pressure within 3 mmHg Pulse accuracy 5 of the read out value Power source 4 AAA batteries Battery lifetime Approx 920 measurements Additional c...

Page 31: ...actured to meet the European and United States standards for non invasive blood pressure monitors EN 1060 3 4 IEC 60601 1 IEC 60601 1 2 EMC IEC 80601 2 30 IEC 60601 1 11 This device complies with the requirements of the Medical Device Directive 93 42 EEC Technical alterations reserved This device is tested according to the ESH protocol and ISO81060 2 2013 Warranty Quality Standards How to Contact ...

Page 32: ...hen China Microlife UAB P Lukšio g 32 08222 Vilnius Lithuania Distributed by ADC 55 Commerce Drive Hauppauge NY 11788 U S A Inspected in the U S A Made in China tel 631 273 9600 1 800 232 2670 fax 631 273 9659 www adctoday com email info adctoday com IB p n 93 6024N 00 rev 1 06 16 2022 Printed in China 0044 ...